Ganapathy K, Kurup P G G, Murali V, Muthukumaran M, Subramanian S Balaji, Velmurugan J
Department of Radiotherapy, Apollo Specialty Hospital, Chennai, Tamil Nadu, India.
J Med Phys. 2013 Jul;38(3):132-8. doi: 10.4103/0971-6203.116372.
The objective of this study is to check the feasibility of in vivo rectal dose measurement in intensity-modulated radiotherapy (IMRT) and CyberKnife treatments for carcinoma prostate. An in-house pelvis phantom made with bee's wax was used in this study. Two cylindrical bone equivalent materials were used to simulate the femur. Target and other critical structures associated with carcinoma prostate were delineated on the treatment planning images by the radiation oncologist. IMRT treatment plan was generated in Oncentra Master Plan treatment planning system and CyberKnife treatment plan was generated in Multiplan treatment planning system. Dose measurements were carried out in phantom and in patient using Gafchromic EBT3 films. RIT software was used to analyze the dose measured by EBT3 films. The measured doses using EBT3 films were compared with the TPS-calculated dose along the anterior rectal wall at multiple points. From the in-phantom measurements, it is observed that the difference between calculated and measured dose was mostly within 5%, except for a few measurement points. The difference between calculated and measured dose in the in-patient measurements was higher than 5% in regions which were away from the target. Gafchromic EBT3 film is a suitable detector for in vivo rectal dose measurements as it offers the possibility of analyzing the dose at multiple points. In addition, the method of extending this in vivo rectal dose measurement technique as a tool for patient-specific quality assurance check is also analyzed.
本研究的目的是检验在前列腺癌的调强放射治疗(IMRT)和射波刀治疗中进行直肠内剂量测量的可行性。本研究使用了由蜂蜡制成的自制盆腔模体。使用两种圆柱形等效骨材料来模拟股骨。放射肿瘤学家在治疗计划图像上勾画出与前列腺癌相关的靶区和其他关键结构。在Oncentra Master Plan治疗计划系统中生成IMRT治疗计划,在Multiplan治疗计划系统中生成射波刀治疗计划。使用Gafchromic EBT3薄膜在模体和患者体内进行剂量测量。使用RIT软件分析EBT3薄膜测量的剂量。将使用EBT3薄膜测量的剂量与TPS在多个点沿直肠前壁计算的剂量进行比较。从模体内测量结果来看,除了少数测量点外,计算剂量与测量剂量之间的差异大多在5%以内。在患者体内测量中,远离靶区的区域计算剂量与测量剂量之间的差异高于5%。Gafchromic EBT3薄膜是一种适合用于直肠内剂量测量的探测器,因为它提供了在多个点分析剂量的可能性。此外,还分析了将这种直肠内剂量测量技术扩展为针对患者的质量保证检查工具的方法。