Dosimetry for Dose Verification of Total Skin Electron Beam Therapy Using Gafchromic® EBT3 Film Dosimetry.

作者信息

Baba Misba Hamid, Singh Benoy Kumar, Wani Shaqul Qamar

机构信息

Department of Physics, Institute of Applied Sciences and Humanities, GLA University, Mathura, Uttar Pradesh, India.

Department of Radiological Physics and B.E, Sher I Kashmir Institute of Medical Sciences, Srinagar, Jammu and Kashmir, India.

出版信息

J Med Phys. 2022 Oct-Dec;47(4):362-366. doi: 10.4103/jmp.jmp_72_22. Epub 2023 Jan 10.

Abstract

BACKGROUND AND PURPOSE

Total skin electron beam therapy (TSEBT) is an important skin-directed radiotherapeutic procedure done in the treatment of cutaneous T-cell lymphomas, namely, mycosis fungoides (MF). This procedure is usually done at larger source-to-surface distances with the patient standing on a rotatory platform. As the patient has to stand in different positions without any rigid immobilization devices, there are chances that the total skin may not get uniformly irradiated which could lead to nonuniform dose distributions. Therefore, all the necessary arrangements should be made to evaluate the dose for different regions of the skin using suitable dosimeters at the radiotherapy centers offering these treatments. This study aimed to evaluate the consistency between the delivered and planned doses during TSEBT using Gafchromic EBT3 film dosimetry.

MATERIALS AND METHODS

The surface dose for the six MF patients treated for TSEBT at our hospital from 2018 to 2022 was measured and evaluated. 2 cm × 2 cm Gafchromic EBT3 films were used to measure skin dose at reference body positions of clinical interest. All the patients were treated with the modified Stanford technique. The irradiated film strips were analyzed for the dose using the IMRT OmniPro software. The doses at respective positions were expressed as mean dose ± standard deviation and the deviation was calculated as the percentage of the prescribed dose.

RESULTS

One hundred and fifty-four Gafchromic EBT3 film strips irradiated on six TSEBT patients showed a maximum dose variation of 2.00 ± 0.14 Gy, in the central body regions. The dose variation in the peripheral areas such as hands and ears was larger. A variation of 2 ± 0.32 Gy was observed on the hands and ears. The uniformity of the dose delivered to maximum body parts was within -7% and +16% for the peripheral areas like hands. The American Association of Physicists in Medicine recommends a dose uniformity of 8% and 4% in the vertical and horizontal patient plane for direct incident beam; however, for oblique incidences like in the modified Stanford technique, the dose variation is about 15%.

CONCLUSION

dosimetry using Gafchromic EBT3 film dosimetry for TSEBT yields objective data to find the under or overdose regions. That can be useful to provide quality treatment, especially when treatments tend to be as complex as TSEBT.

摘要

背景与目的

全身皮肤电子束治疗(TSEBT)是一种重要的针对皮肤的放射治疗方法,用于治疗皮肤T细胞淋巴瘤,即蕈样肉芽肿(MF)。该治疗通常在较大的源皮距下进行,患者站在旋转平台上。由于患者必须在没有任何刚性固定装置的情况下处于不同位置,存在全身皮肤可能无法得到均匀照射的情况,这可能导致剂量分布不均匀。因此,在提供这些治疗的放疗中心,应做出所有必要安排,使用合适的剂量仪评估皮肤不同区域的剂量。本研究旨在使用Gafchromic EBT3薄膜剂量测定法评估TSEBT期间实际给予剂量与计划剂量之间的一致性。

材料与方法

对2018年至2022年在我院接受TSEBT治疗的6例MF患者的表面剂量进行测量和评估。使用2 cm×2 cm的Gafchromic EBT3薄膜在临床关注的参考身体位置测量皮肤剂量。所有患者均采用改良斯坦福技术进行治疗。使用IMRT OmniPro软件分析照射后的薄膜条带的剂量。各个位置的剂量表示为平均剂量±标准差,偏差计算为处方剂量的百分比。

结果

在6例接受TSEBT治疗的患者身上照射的154条Gafchromic EBT3薄膜条带显示,身体中心区域的最大剂量变化为2.00±0.14 Gy。手部和耳部等周边区域的剂量变化较大。手部和耳部观察到的变化为2±0.32 Gy。对于手部等周边区域,给予身体大部分区域的剂量均匀性在-7%至+16%之间。美国医学物理学家协会建议,对于直接入射束,在患者垂直和水平平面上的剂量均匀性分别为8%和4%;然而,对于改良斯坦福技术中的斜入射,剂量变化约为15%。

结论

使用Gafchromic EBT3薄膜剂量测定法进行TSEBT剂量测定可产生客观数据,以发现剂量不足或过量区域。这对于提供高质量治疗可能有用,尤其是当治疗像TSEBT一样复杂时。

https://cdn.ncbi.nlm.nih.gov/pmc/blobs/e42c/9997533/6ee1869ff371/JMP-47-362-g001.jpg