Suppr超能文献

用于测定绵羊血液L-乳酸浓度的Lactate Scout即时检测仪器的可靠性。

Reliability of the Lactate Scout point-of-care instrument for the determination of blood L-lactate concentration in sheep.

作者信息

Kaynar Ozgur, Karapinar Tolga, Hayirli Armagan, Baydar Ersoy

机构信息

Department of Biochemistry, Faculty of Veterinary Medicine, Ataturk University, Erzurum, Turkey.

Department of Internal Medicine, Faculty of Veterinary Medicine, Firat University, Elazig, Turkey.

出版信息

Vet Clin Pathol. 2015 Dec;44(4):559-63. doi: 10.1111/vcp.12288. Epub 2015 Sep 8.

Abstract

BACKGROUND

Data on accuracy and precision of the Lactate Scout point-of-care (POC) analyzer in ovine medicine are lacking.

OBJECTIVE

The purpose of the study was to assess the reliability of the Lactate Scout in sheep.

MATERIALS AND METHODS

Fifty-seven sheep at varying ages with various diseases were included. Blood lactate concentration in samples collected from the jugular vein was measured immediately on the Lactate Scout. Plasma L-lactate concentration was measured by the Cobas autoanalyzer as the reference method. Data were subjected to Student's t-test, Passing-Bablok regression, and Bland-Altman plot analyses for comparison and assessment of accuracy, precision, and reliability.

RESULTS

Plasma l-lactate concentration was consistently lower than blood L-lactate concentration (3.06 ± 0.24 vs 3.3 ± 0.3 mmol/L, P < .0001). There was a positive correlation between plasma and blood L-lactate concentrations (r = .98, P < .0001). The Lactate Scout had 99% accuracy and 98% precision with the reference method. Blood (Y) and plasma (X) L-lactate concentrations were fitted to Y = 0.28 + 1.00 · X, with a residual standard deviation of 0.31 and a negligible deviation from the identity line (P = .93). The bias was fitted to Y = 0.10 + 0.05 · X, with Sy.x of 0.44 (P < .07).

CONCLUSIONS

The Lactate Scout has high accuracy and precision, with a negligible bias. It is a reliable POC analyzer to assess L-lactate concentration in ovine medicine.

摘要

背景

缺乏关于乳酸 Scout 即时检验(POC)分析仪在羊医学中准确性和精密度的数据。

目的

本研究的目的是评估乳酸 Scout 在绵羊中的可靠性。

材料与方法

纳入了 57 只不同年龄、患有各种疾病的绵羊。从颈静脉采集的样本中的血乳酸浓度立即用乳酸 Scout 进行测量。血浆 L-乳酸浓度通过 Cobas 自动分析仪测量作为参考方法。对数据进行 Student's t 检验、Passing-Bablok 回归分析和 Bland-Altman 图分析,以比较和评估准确性、精密度和可靠性。

结果

血浆 L-乳酸浓度始终低于血 L-乳酸浓度(3.06±0.24 对 3.3±0.3 mmol/L,P<.0001)。血浆和血 L-乳酸浓度之间存在正相关(r =.98,P<.0001)。乳酸 Scout 与参考方法相比具有 99%的准确性和 98%的精密度。血(Y)和血浆(X)L-乳酸浓度拟合为 Y = 0.28 + 1.00·X,残余标准差为 0.31,与恒等线的偏差可忽略不计(P =.93)。偏差拟合为 Y = 0.10 + 0.05·X,Sy.x 为 0.44(P<.07)。

结论

乳酸 Scout 具有高准确性和精密度,偏差可忽略不计。它是评估羊医学中 L-乳酸浓度的可靠 POC 分析仪。