To summarize the approaches used to manage exposure of patients to inadequately sterilized neurosurgical instruments contaminated as a result of Creutzfeldt-Jakob disease (CJD).
Information on past CJD exposure incidents reported to the Centers for Disease Control and Prevention (CDC) was aggregated and summarized. In addition, inactivation studies were reviewed, and data from selected publications were provided for reference.
Nineteen incidents of patient exposure to potentially CJD-contaminated instruments were reported to the CDC, including 17 that involved intracranial procedures and 2 that involved ophthalmologic procedures. In more than 50% of incidents, the neurosurgical procedures were performed for diagnostic work up of the index patients. At least 12 of the hospitals had multiple neurosurgical sets, and the CJD-contaminated instruments could not be identified in 11 of 19 hospitals. In 12 of 15 hospitals with neurosurgical incidents, a decision was made to notify patients of their potential exposure.
Neurosurgical instruments used for treatment of patients with suspected or diagnosed CJD or patients whose diagnosis is unclear should be promptly identified and sterilized using recommended CJD decontamination protocols. Inability to trace instruments complicates appropriate management of exposure incidents. The feasibility of instituting instrument tracking procedures should be considered.
总结处理因克雅氏病(CJD)而导致受污染的不充分消毒神经外科器械暴露于患者的方法。
汇总并总结了向疾病控制与预防中心(CDC)报告的过去 CJD 暴露事件的信息。此外,还回顾了失活研究,并提供了部分已发表文献的数据供参考。
CDC 共报告了 19 例患者接触潜在 CJD 污染器械的事件,其中 17 例涉及颅内手术,2 例涉及眼科手术。在超过 50%的事件中,神经外科手术是为索引患者的诊断性检查而进行的。在 19 家医院中有 12 家医院有多个神经外科器械包,在 19 家医院中有 11 家无法确定被 CJD 污染的器械。在有神经外科事件的 15 家医院中的 12 家,决定通知患者他们可能接触到了这些器械。
用于治疗疑似或确诊 CJD 患者或诊断不明的患者的神经外科器械应立即识别,并按照推荐的 CJD 去污方案进行消毒。无法追踪器械会使暴露事件的适当处理变得复杂。应考虑制定器械追踪程序的可行性。