Ploem J S, van Driel-Kulker A M, Goyarts-Veldstra L, Ploem-Zaaijer J J, Verwoerd N P, van der Zwan M
Histochemistry. 1986;84(4-6):549-55. doi: 10.1007/BF00482990.
This paper describes the application of image analysis combined with a quantitative staining method for the analysis of cervical specimens. The image analysis is carried out with the Leyden Television Analysis System, LEYTAS, of which two versions are described. LEYTAS-1 as well as LEYTAS-2 have both been designed with a high degree of flexibility and interaction facilities. A much wider range of image analysis programs is however, possible with LEYTAS-2, enabling many applications. LEYTAS-1, the earlier version, consists of a Leitz microscope with automated functions, a TV camera, the Texture Analysis System (TAS, Leitz), a four-bit grey value memory and a minicomputer (PDP 11/23). Using this instrumentation 1,500 cervical smears prepared from cell suspensions and stained with acriflavin-Feulgen-Sits have been analysed in a completely automated procedure. Image transformations working in parallel on entire fields, have been used for cell selection and artefact rejection. Resulting alarms, consisting of selected single cells and non-rejected artefacts are stored in the grey value memory, which is displayed on a TV monitor. This option allows visual interaction after the machine diagnosis has been made. The machine diagnosis was correct in 320 out 321 specimens with a severe dysplasia or more serious lesion. The false positive rate in 561 morphologically negative specimens (normal and inflammation) was 16% (machine diagnosis). Visual interaction by subtracting the visually recognized false alarms from the total number of alarms reduces the false positive rate to 11%.(ABSTRACT TRUNCATED AT 250 WORDS)
本文描述了图像分析结合定量染色方法在宫颈标本分析中的应用。图像分析使用莱顿电视分析系统(LEYTAS)进行,文中介绍了该系统的两个版本。LEYTAS - 1和LEYTAS - 2都设计有高度的灵活性和交互功能。然而,LEYTAS - 2具备更广泛的图像分析程序,适用于多种应用。早期版本的LEYTAS - 1由一台具有自动功能的徕卡显微镜、一台电视摄像机、纹理分析系统(TAS, 徕卡)、一个四位灰度值存储器和一台小型计算机(PDP 11/23)组成。使用这套仪器,对1500份由细胞悬液制备并用吖啶黄素 - 福尔根 - 西茨染色的宫颈涂片进行了完全自动化分析。在整个视野上并行运行的图像变换用于细胞选择和伪像排除。产生的警报,包括选定的单个细胞和未排除的伪像,存储在灰度值存储器中,并显示在电视监视器上。此选项允许在进行机器诊断后进行视觉交互。在321份有严重发育异常或更严重病变的标本中,机器诊断正确的有320份。在561份形态学阴性标本(正常和炎症)中,假阳性率为16%(机器诊断)。通过从警报总数中减去视觉识别出的误报进行视觉交互,可将假阳性率降至11%。(摘要截取自250字)