血糖仪相对于行业参考标准的有效性和可靠性。

Validity and Reliability of a Glucometer Against Industry Reference Standards.

作者信息

Salacinski Amanda J, Alford Micah, Drevets Kathryn, Hart Sarah, Hunt Brian E

机构信息

Department of Kinesiology and Physical Education, Northern Illinois University, DeKalb, IL, USA

Department of Applied Health Science, Wheaton College, Wheaton, IL, USA.

出版信息

J Diabetes Sci Technol. 2014 Jan;8(1):95-99. doi: 10.1177/1932296813514315. Epub 2014 Jan 1.

Abstract

As an appealing alternative to reference glucose analyzers, portable glucometers are recommended for self-monitoring at home, in the field, and in research settings. The purpose was to characterize the accuracy and precision, and bias of glucometers in biomedical research. Fifteen young (20-36 years; mean = 24.5), moderately to highly active men (n = 10) and women (n = 5), defined by exercising 2 to 3 times a week for the past 6 months, were given an oral glucose tolerance test (OGTT) after an overnight fast. Participants ingested 50, 75, or 150 grams of glucose over a 5-minute period. The glucometer was compared to a reference instrument. The glucometer had 39% of values within 15% of measurements made using the reference instrument ranging from 45.05 to 169.37 mg/dl. There was both a proportional (-0.45 to -0.39) and small fixed (5.06 and 0.90 mg/dl) bias. Results of the present study suggest that the glucometer provided poor validity and reliability results compared to the results provided by the reference laboratory analyzer. The portable glucometers should be used for patient management, but not for diagnosis, treatment, or research purposes.

摘要

作为参考血糖仪的一种有吸引力的替代方案,便携式血糖仪被推荐用于家庭、野外和研究环境中的自我监测。目的是在生物医学研究中表征血糖仪的准确性、精密度和偏差。15名年轻(20 - 36岁;平均 = 24.5岁)、中度至高度活跃的男性(n = 10)和女性(n = 5),他们在过去6个月每周锻炼2至3次,在空腹过夜后进行口服葡萄糖耐量试验(OGTT)。参与者在5分钟内摄入50、75或150克葡萄糖。将血糖仪与参考仪器进行比较。血糖仪在使用参考仪器测量值的15%范围内的数值占39%,测量值范围为45.05至169.37毫克/分升。存在比例偏差(-0.45至-0.39)和小的固定偏差(5.06和0.90毫克/分升)。本研究结果表明,与参考实验室分析仪提供的结果相比,血糖仪的有效性和可靠性结果较差。便携式血糖仪应用于患者管理,但不适用于诊断、治疗或研究目的。

相似文献

5
8
Glucometers-need for external quality control.血糖仪——外部质量控制的必要性。
Indian J Clin Biochem. 2000 Jul;15(2):128-33. doi: 10.1007/BF02883741.

引用本文的文献

3
Comparative Study Between Blood Glucose Meters in a Maternity Center.某妇产中心血糖仪的对比研究
Womens Health Rep (New Rochelle). 2024 Jul 26;5(1):572-578. doi: 10.1089/whr.2024.0030. eCollection 2024.
5
[Not Available].[无可用内容]。
Adv Lab Med. 2024 Feb 19;5(2):197-204. doi: 10.1515/almed-2024-0017. eCollection 2024 Jun.

本文引用的文献

1
Accuracy of the CONTOUR® blood glucose monitoring system.CONTOUR®血糖监测系统的准确性。
J Diabetes Sci Technol. 2011 Jul 1;5(4):1009-13. doi: 10.1177/193229681100500425.
4
Variability among five over-the-counter blood glucose monitors.五种非处方血糖监测仪之间的差异。
Clin Chim Acta. 2006 Feb;364(1-2):292-7. doi: 10.1016/j.cca.2005.07.027. Epub 2005 Sep 6.
8
Fundamentals of resistance training: progression and exercise prescription.抗阻训练基础:进展与运动处方
Med Sci Sports Exerc. 2004 Apr;36(4):674-88. doi: 10.1249/01.mss.0000121945.36635.61.
9
Evolution of analytical performance in portable glucose meters in the last decade.
Diabetes Care. 2003 Apr;26(4):1170-5. doi: 10.2337/diacare.26.4.1170.
10
Tests of glycemia in diabetes.糖尿病血糖检测
Diabetes Care. 2003 Jan;26 Suppl 1:S106-8. doi: 10.2337/diacare.26.2007.s106.

文献检索

告别复杂PubMed语法,用中文像聊天一样搜索,搜遍4000万医学文献。AI智能推荐,让科研检索更轻松。

立即免费搜索

文件翻译

保留排版,准确专业,支持PDF/Word/PPT等文件格式,支持 12+语言互译。

免费翻译文档

深度研究

AI帮你快速写综述,25分钟生成高质量综述,智能提取关键信息,辅助科研写作。

立即免费体验