使用英国高血压协会(BHS)和医疗仪器促进协会(AAMI)协议指南对Datascope Accutorr Plus™进行验证。

Validation of Datascope Accutorr Plus™ using British Hypertension Society (BHS) and Association for the Advancement of Medical Instrumentation (AAMI) protocol guidelines.

作者信息

Khawaja Rajab Ali, Qureshi Riaz, Mansure Abdul Hay, Yahya Mohammad Emran

机构信息

Department of Family and Community Medicine, King Khalid University Hospital, King Saud University, Riyadh, Saudi Arabia.

出版信息

J Saudi Heart Assoc. 2010 Jan;22(1):1-5. doi: 10.1016/j.jsha.2010.03.001. Epub 2010 Mar 10.

Abstract

OBJECTIVE

To assure the accuracy and reliability of blood pressure measurement by non-invasive blood pressure monitor using Datascope Accutorr Plus™ (Paramus, NJ, USA) against mercury manometer, among adult male participants.

METHOD

Eighty participants from a family physician's office at a teaching hospital were recruited. One hundred and sixty measurements of blood pressure were performed according to BHS technique protocol.

STATISTICAL ANALYSIS

Descriptive analysis was done according to the AAMI and BHS protocol guidelines. The limits of agreement between the device and the standard were plotted using the method of Bland and Altman plot.

RESULTS

The mean difference ± SD between the Datascope Acutorr Plus™ and observer was 2.7 ± 5.2 mm Hg and 1.5 ± 3.26 mm Hg for systolic and diastolic blood pressure, respectively. Datascope Acutorr Plus™ obtained A/A grading for both systolic and diastolic blood pressure.

CONCLUSION

Datascope Accutorr Plus™ (Paramus, NJ, USA) satisfies BHS and AAMI validation protocols for both systolic and diastolic BP and may be recommended for everyday use for BP monitoring at home and in clinical use for adult population.

摘要

目的

在成年男性参与者中,使用美国新泽西州帕拉默斯市Datascope公司的Accutorr Plus™无创血压监测仪,对比汞柱式血压计,确保血压测量的准确性和可靠性。

方法

招募了一家教学医院家庭医生办公室的80名参与者。按照英国高血压学会(BHS)技术方案进行了160次血压测量。

统计分析

根据美国医疗器械促进协会(AAMI)和BHS方案指南进行描述性分析。使用布兰德-奥特曼图法绘制该设备与标准之间的一致性界限。

结果

Datascope Acutorr Plus™与观察者之间收缩压和舒张压的平均差值±标准差分别为2.7±5.2毫米汞柱和1.5±3.26毫米汞柱。Datascope Acutorr Plus™的收缩压和舒张压均获得A/A级评分。

结论

美国新泽西州帕拉默斯市Datascope公司的Accutorr Plus™满足BHS和AAMI关于收缩压和舒张压的验证方案,可推荐用于家庭日常血压监测以及成年人群的临床使用。

相似文献

3
Clinical evaluation of the Trimline blood pressure cuffs with the Accutorr Plus Monitor.
Blood Press Monit. 2003 Jun;8(3):137-40. doi: 10.1097/00126097-200306000-00008.

引用本文的文献

1
Family history of dementia and brain health in childhood and middle age: a prospective community-based study.
Eur J Epidemiol. 2024 Oct;39(10):1151-1160. doi: 10.1007/s10654-024-01160-2. Epub 2024 Oct 10.
2
Relationship of Sodium Intake With Granulocytes, Renal and Cardiovascular Outcomes in the Prospective EPIC-Norfolk Cohort.
J Am Heart Assoc. 2022 Jul 5;11(13):e023727. doi: 10.1161/JAHA.121.023727. Epub 2022 Jun 22.
4
Design and overview of the Origins of Alzheimer's Disease Across the Life course (ORACLE) study.
Eur J Epidemiol. 2021 Jan;36(1):117-127. doi: 10.1007/s10654-020-00696-3. Epub 2020 Dec 16.
5
Maternal cardiovascular adaptation to twin pregnancy: a population-based prospective cohort study.
BMC Pregnancy Childbirth. 2020 May 29;20(1):327. doi: 10.1186/s12884-020-02994-w.
6
Early Pregnancy Cardiovascular Health and Subclinical Atherosclerosis.
J Am Heart Assoc. 2019 Aug 6;8(15):e011394. doi: 10.1161/JAHA.118.011394. Epub 2019 Jul 23.
7
Cardiovascular Risk Factors Track From Mother to Child.
J Am Heart Assoc. 2018 Oct 2;7(19):e009536. doi: 10.1161/JAHA.118.009536.
9
Nonadherence to antihypertensive medications and associated factors in general medicine clinics.
Patient Prefer Adherence. 2016 Aug 1;10:1415-9. doi: 10.2147/PPA.S100735. eCollection 2016.
10

本文引用的文献

3
Accuracy of the HoMedics BPA-300, a home blood pressure monitor using the auscultatory method.
Blood Press Monit. 2007 Oct;12(5):339-43. doi: 10.1097/MBP.0b013e3282c9ac12.
5
Validation of the DINAMAP ProCare blood pressure device according to the international protocol in an adult population.
Blood Press Monit. 2006 Oct;11(5):293-6. doi: 10.1097/01.mbp.0000217998.96967.fb.
6
Validation of three oscillometric blood pressure devices against auscultatory mercury sphygmomanometer in children.
Blood Press Monit. 2006 Oct;11(5):281-91. doi: 10.1097/01.mbp.0000209082.09623.b4.
10
Accuracy of the Dinamap 1846 XT automated blood pressure monitor.
J Hum Hypertens. 2002 Sep;16(9):647-52. doi: 10.1038/sj.jhh.1001463.

文献AI研究员

20分钟写一篇综述,助力文献阅读效率提升50倍。

立即体验

用中文搜PubMed

大模型驱动的PubMed中文搜索引擎

马上搜索

文档翻译

学术文献翻译模型,支持多种主流文档格式。

立即体验