The examination of the physical properties of the instrument showed several possibilities of meter-error mainly due to constructive disadvantages. After having found this a special experimental arrangement was worked out to exclude errors. Then good linear correlations of the meter-reading and the analytical DPD-Method for free Chlorine were found for pH-values ranging from 3.7 to 11. Further the results of a test under practical conditions in a swimming pool are shown and the possibility to discriminate different types of waters by their chlorine demand under constant-titration.
对该仪器物理特性的检测表明,主要由于结构缺陷,存在几种仪表误差的可能性。发现这一情况后,设计了一种特殊的实验装置以排除误差。然后,发现对于pH值在3.7至11范围内,仪表读数与游离氯的分析DPD方法之间具有良好的线性相关性。此外,还展示了在游泳池实际条件下的测试结果,以及在恒定量滴定条件下通过氯需求量区分不同类型水的可能性。