Suppr超能文献

使用iSTAT和Piccolo分析仪进行即时检测与固定医院化学分析平台对非血清样本进行体液分析。

Bodily fluid analysis of non-serum samples using point-of-care testing with iSTAT and Piccolo analyzers versus a fixed hospital chemistry analytical platform.

作者信息

Londeree William, Davis Konrad, Helman Donald, Abadie Jude

机构信息

Department of Medicine, Tripler Army Medical Center, Honolulu, HI (WL).

出版信息

Hawaii J Med Public Health. 2014 Sep;73(9 Suppl 1):3-8.

Abstract

INTRODUCTION

Forward deployed military medical units can provide sophisticated medical care with limited resources. Point-of-Care Testing (POCT) may facilitate care and expedite diagnosis. This study assessed the accuracy of results for POCT for non-serum samples (pleural, peritoneal, and cerebrospinal fluid) using iSTAT and Piccolo hand-held devices compared with results obtained using a hospital chemistry analyzer.

METHODS

Pleural, peritoneal, and cerebrospinal fluids obtained during routine care were simultaneously analyzed on a Vitros 5600 automated clinical chemistry hospital analyzer, iSTAT, and Piccolo POCT devices.

RESULTS

POCT results were highly correlated with the Vitros 5600 for pleural fluid LDH, glucose, and triglycerides (TG); for peritoneal fluid bilirubin, TG, glucose, albumin, and protein; and glucose for cerebrospinal fluid.

CONCLUSION

POCT results for non-serum samples from pleural, peritoneal, and cerebrospinal fluid correlate with standard hospital chemistry analysis. The results of this study demonstrate potential for possible new diagnostic roles for POCT in resource-limited environments.

摘要

引言

前沿部署的军事医疗单位能够在资源有限的情况下提供复杂的医疗服务。即时检验(POCT)可能有助于医疗服务并加快诊断。本研究评估了使用iSTAT和Piccolo手持式设备对非血清样本(胸腔积液、腹腔积液和脑脊液)进行POCT检测结果的准确性,并与使用医院化学分析仪获得的结果进行比较。

方法

在常规护理过程中获取的胸腔积液、腹腔积液和脑脊液同时在Vitros 5600自动化临床化学医院分析仪、iSTAT和Piccolo POCT设备上进行分析。

结果

POCT检测结果与Vitros 5600对胸腔积液乳酸脱氢酶(LDH)、葡萄糖和甘油三酯(TG);腹腔积液胆红素、TG、葡萄糖、白蛋白和蛋白质;以及脑脊液葡萄糖的检测结果高度相关。

结论

胸腔积液、腹腔积液和脑脊液非血清样本的POCT检测结果与标准医院化学分析结果相关。本研究结果表明POCT在资源有限环境中可能具有新的诊断作用。

相似文献

2
Peritoneal and Pleural Fluid Chemistry Measurements Performed on Three Chemistry Platforms.
Lab Med. 2019 Apr 8;50(2):145-149. doi: 10.1093/labmed/lmy056.
4
Body Fluid Testing at John F. Kennedy Medical Center in Liberia.
Am J Clin Pathol. 2019 Jun 5;152(1):86-90. doi: 10.1093/ajcp/aqz027.
5
Assessment of liver function tests on Piccolo Xpress point of care chemistry analyzer in a pediatric hospital.
Pract Lab Med. 2015 Oct 3;3:1-7. doi: 10.1016/j.plabm.2015.09.002. eCollection 2015 Dec 1.
6
Performance specifications of common chemistry analytes on the AU series of chemistry analyzers for miscellaneous body fluids.
Clin Chim Acta. 2013 Nov 15;426:121-6. doi: 10.1016/j.cca.2013.08.011. Epub 2013 Aug 24.
7
ASVCP guidelines: quality assurance for point-of-care testing in veterinary medicine.
Vet Clin Pathol. 2013 Dec;42(4):405-23. doi: 10.1111/vcp.12099.
9
Lactate point-of-care testing for acidosis: Cross-comparison of two devices with routine laboratory results.
Pract Lab Med. 2015 Dec 24;4:41-49. doi: 10.1016/j.plabm.2015.12.005. eCollection 2016 Apr 1.
10
Compact dry chemistry instruments.
Southeast Asian J Trop Med Public Health. 1999;30 Suppl 3:122-9.

引用本文的文献

1
Immunological Disorders: Gradations and the Current Approach in Laboratory Diagnostics.
Pathophysiology. 2025 Apr 18;32(2):17. doi: 10.3390/pathophysiology32020017.
2
A dual-color plasmonic immunosensor for salivary cortisol measurement.
Nanoscale Adv. 2022 Nov 8;5(2):329-336. doi: 10.1039/d2na00563h. eCollection 2023 Jan 18.
3
Wearable sensors enable personalized predictions of clinical laboratory measurements.
Nat Med. 2021 Jun;27(6):1105-1112. doi: 10.1038/s41591-021-01339-0. Epub 2021 May 24.

本文引用的文献

1
How point-of-care testing could drive innovation in global health.
N Engl J Med. 2013 Jun 13;368(24):2319-24. doi: 10.1056/NEJMsb1214197.
2
Analytical performance of three point of care methods for pleural fluid pH analysis.
Clin Biochem. 2013 Aug;46(12):1139-1141. doi: 10.1016/j.clinbiochem.2013.04.021. Epub 2013 Apr 30.
4
Chylothorax and cholesterol pleural effusion.
Semin Respir Crit Care Med. 2010 Dec;31(6):743-50. doi: 10.1055/s-0030-1269834. Epub 2011 Jan 6.
6
Global health funding: how much, where it comes from and where it goes.
Health Policy Plan. 2009 Nov;24(6):407-17. doi: 10.1093/heapol/czp026. Epub 2009 Jul 1.
7
Factors influencing the measurement of pleural fluid pH.
Curr Opin Pulm Med. 2009 Jul;15(4):353-7. doi: 10.1097/MCP.0b013e32832b98d4.
8
Clinically important factors influencing the diagnostic measurement of pleural fluid pH and glucose.
Am J Respir Crit Care Med. 2008 Sep 1;178(5):483-90. doi: 10.1164/rccm.200801-062OC. Epub 2008 Jun 12.
10
Executive summary. The National Academy of Clinical Biochemistry Laboratory Medicine Practice Guideline: evidence-based practice for point-of-care testing.
Clin Chim Acta. 2007 Apr;379(1-2):14-28; discussion 29-30. doi: 10.1016/j.cca.2006.12.025. Epub 2007 Jan 12.

文献AI研究员

20分钟写一篇综述,助力文献阅读效率提升50倍。

立即体验

用中文搜PubMed

大模型驱动的PubMed中文搜索引擎

马上搜索

文档翻译

学术文献翻译模型,支持多种主流文档格式。

立即体验