Sponsel W E, Kaufman P L, Strinden T I, DePaul K L, Bowes H N, Olander K W, Barnebey H S
Department of Ophthalmology, University of Wisconsin Medical School, Madison.
Acta Ophthalmol (Copenh). 1989 Oct;67(5):567-72. doi: 10.1111/j.1755-3768.1989.tb04109.x.
The recently-introduced Pulsair non-contact tonometer (Keeler Instruments, Inc, Broomall, PA) was evaluated against the Goldmann applanation tonometer in cannulated post mortem human eyes, and in living subjects at three clinical centers. Statistical analysis of the findings revealed strong linear relationships between Goldmann or manometric and Pulsair intraocular pressure readings, with correlation coefficients ranging from 0.79 to 0.97. However, with the Pulsair reading as the dependent variable, the slope of the relationship in each case was significantly less than 1.0 (range 0.65 to 0.83), while the intercept was positive (range 1.31 to 2.33 mmHg). Thus, the Pulsair tended to read low at intraocular pressure above the normal range. Conversion factors and a calibration table allowing intraocular pressure estimates more comparable to Goldmann tonometry using the current (J-series) Pulsair are provided herein. A calibration algorithm could be integrated into the circuitry of future production models of the Pulsair instrument.
在三个临床中心,针对插管的人死后眼球以及活体受试者,对最近推出的Pulsair非接触眼压计(Keeler Instruments, Inc, Broomall, PA)与戈德曼压平眼压计进行了评估。对研究结果的统计分析显示,戈德曼眼压计测量值或压力测量值与Pulsair眼压计测量值之间存在很强的线性关系,相关系数在0.79至0.97之间。然而,以Pulsair测量值作为因变量时,每种情况下关系的斜率均显著小于1.0(范围为0.65至0.83),而截距为正值(范围为1.31至2.33 mmHg)。因此,当眼压高于正常范围时,Pulsair眼压计的读数往往偏低。本文提供了换算因子和校准表,以便使用当前(J系列)Pulsair眼压计得出更接近戈德曼眼压测量法的眼压估计值。校准算法可集成到Pulsair仪器未来生产型号的电路中。