Anodisation Increases Integration of Unloaded Titanium Implants in Sheep Mandible.

作者信息

Duncan Warwick J, Lee Min-Ho, Bae Tae-Sung, Lee Sook-Jeong, Gay Jennifer, Loch Carolina

机构信息

Sir John Walsh Research Institute, Faculty of Dentistry, University of Otago, Dunedin, New Zealand.

Department of Dental Biomaterials and Institute of Biodegradable Materials, Institute of Oral Bioscience and BK21 Plus Project, School of Dentistry, Chonbuk National University, Jeonju, Republic of Korea.

出版信息

Biomed Res Int. 2015;2015:857969. doi: 10.1155/2015/857969. Epub 2015 Sep 8.

Abstract

Spark discharge anodic oxidation forms porous TiO2 films on titanium implant surfaces. This increases surface roughness and concentration of calcium and phosphate ions and may enhance early osseointegration. To test this, forty 3.75 mm × 13 mm titanium implants (Megagen, Korea) were placed into healed mandibular postextraction ridges of 10 sheep. There were 10 implants per group: RBM surface (control), RBM + anodised, RBM + anodised + fluoride, and titanium alloy + anodised surface. Resonant frequency analysis (RFA) was measured in implant stability quotient (ISQ) at surgery and at sacrifice after 1-month unloaded healing. Mean bone-implant contact (% BIC) was measured in undemineralised ground sections for the best three consecutive threads. One of 40 implants showed evidence of failure. RFA differed between groups at surgery but not after 1 month. RFA values increased nonsignificantly for all implants after 1 month, except for controls. There was a marked difference in BIC after 1-month healing, with higher values for alloy implants, followed by anodised + fluoride and anodised implants. Anodisation increased early osseointegration of rough-surfaced implants by 50-80%. RFA testing lacked sufficient resolution to detect this improvement. Whether this gain in early bone-implant contact is clinically significant is the subject of future experiments.

摘要

火花放电阳极氧化在钛种植体表面形成多孔二氧化钛薄膜。这增加了表面粗糙度以及钙和磷酸根离子的浓度,并可能增强早期骨整合。为了对此进行测试,将40颗3.75毫米×13毫米的钛种植体(韩国美格真公司)植入10只绵羊愈合的下颌拔牙后牙槽嵴中。每组有10颗种植体:RBM表面(对照)、RBM + 阳极氧化、RBM + 阳极氧化 + 氟化物以及钛合金 + 阳极氧化表面。在手术时以及1个月无负载愈合后的处死时,通过种植体稳定性商数(ISQ)测量共振频率分析(RFA)。在未脱钙的磨片上测量连续最佳的三个螺纹的平均骨 - 种植体接触率(%BIC)。40颗种植体中有1颗显示出失败迹象。手术时不同组之间的RFA存在差异,但1个月后无差异。除对照组外,所有种植体在1个月后RFA值均无显著增加。1个月愈合后BIC存在显著差异,合金种植体的值更高,其次是阳极氧化 + 氟化物和阳极氧化种植体。阳极氧化使粗糙表面种植体的早期骨整合提高了50 - 80%。RFA测试缺乏足够的分辨率来检测这种改善。早期骨 - 种植体接触的这种增加在临床上是否具有显著意义是未来实验的主题。

https://cdn.ncbi.nlm.nih.gov/pmc/blobs/e331/4578837/1390105bd8dd/BMRI2015-857969.001.jpg

文献AI研究员

20分钟写一篇综述,助力文献阅读效率提升50倍。

立即体验

用中文搜PubMed

大模型驱动的PubMed中文搜索引擎

马上搜索

文档翻译

学术文献翻译模型,支持多种主流文档格式。

立即体验