Anderson J D, Trombley C, Cimolai N
Division of Medical Microbiology, British Columbia's Children's Hospital, Vancouver, Canada.
J Clin Microbiol. 1989 Apr;27(4):721-3. doi: 10.1128/jcm.27.4.721-723.1989.
The sensitivity of the BACTEC NR660 blood culture system was assessed by using paired bottles of anaerobic (NR7A) and resin-containing aerobic (NR16A) media and conditions and organisms which simulated those found in pediatric practice. Corresponding media (7D and 16B) of the established BACTEC 460 radiometric method served as controls. The performances of the two systems were similar with 50 isolates of 10 aerobic organisms (aerobic medium) and with 21 isolates of 15 strict anaerobes (anaerobic medium).
通过使用成对的厌氧(NR7A)和含树脂需氧(NR16A)培养基以及模拟儿科实际情况的条件和生物体,对BACTEC NR660血培养系统的敏感性进行了评估。既定的BACTEC 460放射测量法的相应培养基(7D和16B)用作对照。两种系统对10种需氧生物体的50个分离株(需氧培养基)以及15种严格厌氧菌的21个分离株(厌氧培养基)的检测性能相似。