Brannon P, Kiehn T E
J Clin Microbiol. 1986 Nov;24(5):886-7. doi: 10.1128/jcm.24.5.886-887.1986.
The Isolator 10 lysis-centrifugation blood culture system (E. I. du Pont de Nemours & Co., Inc., Wilmington, Del.) and the BACTEC 16B-17D radiometric resin system (Johnston Laboratories, Inc., Towson, Md.) both remove antimicrobial agents from the blood for culture. We compared these two systems for recovery of aerobic bacteria, facultatively anaerobic bacteria, and yeasts. A total of 5,000 blood cultures yielded 467 clinically significant isolates. Both systems recovered 350 (75%) organisms, 56 (12%) were detected by Isolator only, and 61 (13%) were detected by BACTEC resin bottles only. No group of organisms was isolated significantly more often from either system.
Isolator 10裂解离心血液培养系统(E. I. 杜邦公司,特拉华州威尔明顿)和BACTEC 16B - 17D放射性树脂系统(约翰斯顿实验室公司,马里兰州陶森)均可从血液中去除抗菌剂以进行培养。我们比较了这两种系统对需氧菌、兼性厌氧菌和酵母菌的回收率。总共5000份血液培养物产生了467株具有临床意义的分离株。两种系统均回收了350株(75%)微生物,仅Isolator检测到56株(12%),仅BACTEC树脂瓶检测到61株(13%)。没有哪一组微生物从任一系统中分离出来的频率明显更高。