Tejwani Sushma, Dinakaran Shoruba, Joshi Anuja, Shetty Rohit, Sinha Roy Abhijit
Department of Glaucoma, Narayana Nethralaya, Bengaluru, Karnataka, India.
Indian J Ophthalmol. 2015 Nov;63(11):815-20. doi: 10.4103/0301-4738.171956.
To study the correlation and effect of sequential measurement of intraocular pressure (IOP) with Goldmann applanation tonometer (GAT), ocular response analyzer (ORA), dynamic contour tonometer (DCT), and Corvis ST.
Observational cross-sectional series from the comprehensive clinic of a tertiary eye care center seen during December 2012.
One hundred and twenty-five study eyes of 125 patients with normal IOP and biomechanical properties underwent IOP measurement on GAT, DCT, ORA, and Corvis ST; in four different sequences. Patients with high refractive errors, recent surgeries, glaucoma, and corneal disorders were excluded so as to rule out patients with evident altered corneal biomechanics.
Linear regression and Bland-Altman using MedCalc software.
Multivariate analysis of variance with repeated measures showed no influence of sequence of device use on IOP (P = 0.85). Linear regression r2 between GAT and Corvis ST, Corvis ST and Goldmann-correlated IOP (IOPg), and DCT and Corvis ST were 0.37 (P = 0.675), 0.63 (P = 0.607), and 0.19 (P = 0.708), respectively. The Bland-Altman agreement of Corvis ST with GAT, corneal compensated IOP, and IOPg was 2 mmHg (-5.0 to + 10.3), -0.5 mmHg (-8.1 to 7.1), and 0.5 mmHg (-6.2 to 7.1), respectively. Intraclass correlation coefficient for repeatability ranged from 0.81 to 0.96.
Correlation between Corvis ST and ORA was found to be good and not so with GAT. However, agreement between the devices was statistically insignificant, and no influence of sequence was observed.
研究使用戈德曼压平眼压计(GAT)、眼反应分析仪(ORA)、动态轮廓眼压计(DCT)和Corvis ST对眼压(IOP)进行连续测量的相关性及影响。
2012年12月期间在一家三级眼科护理中心综合诊所进行的观察性横断面系列研究。
对125例眼压和生物力学特性正常的患者的125只研究眼,采用GAT、DCT、ORA和Corvis ST以四种不同顺序测量眼压。排除高度屈光不正、近期手术、青光眼和角膜疾病患者,以排除角膜生物力学明显改变的患者。
使用MedCalc软件进行线性回归和布兰德-奥特曼分析。
重复测量的多变量方差分析显示,设备使用顺序对眼压无影响(P = 0.85)。GAT与Corvis ST、Corvis ST与戈德曼相关眼压(IOPg)以及DCT与Corvis ST之间的线性回归r2分别为0.37(P = 0.675)、0.63(P = 0.607)和0.19(P = 0.708)。Corvis ST与GAT、角膜补偿眼压和IOPg的布兰德-奥特曼一致性分别为2 mmHg(-5.0至+10.3)、-0.5 mmHg(-8.1至7.1)和0.5 mmHg(-6.2至7.1)。重复性的组内相关系数范围为0.81至0.96。
发现Corvis ST与ORA之间相关性良好,与GAT相关性不佳。然而,各设备之间的一致性在统计学上无显著意义,且未观察到顺序的影响。