Evaluation of the technical performance of three different commercial digital breast tomosynthesis systems in the clinical environment.

作者信息

Rodríguez-Ruiz A, Castillo M, Garayoa J, Chevalier M

机构信息

Medical Physics Group, Radiology Department, Medical School, Complutense University of Madrid, Pza. Ramón y Cajal, s/n, 28040 Madrid, Spain.

Medical Physics Group, Radiology Department, Medical School, Complutense University of Madrid, Pza. Ramón y Cajal, s/n, 28040 Madrid, Spain.

出版信息

Phys Med. 2016 Jun;32(6):767-77. doi: 10.1016/j.ejmp.2016.05.001. Epub 2016 May 11.

Abstract

The aim of this work was to research and evaluate the performance of three different digital breast tomosynthesis (DBT) systems in the clinical environment (Siemens Mammomat Inspiration, Hologic Selenia Dimensions, and Fujifilm Amulet Innovality). The characterization included the study of the detector, the automatic exposure control, and the resolution of DBT projections and reconstructed planes. The modulation transfer function (MTF) of the DBT projections was measured with a 1mm thick steel edge, showing a strong anisotropy (30-40% lower MTF0.5 frequencies in the tube travel direction). The in-plane MTF0.5, measured with a 25μm tungsten wire, ranges from 1.3 to 1.8lp/mm in the tube-travel direction and between 2.4 and 3.7lp/mm in the chest wall-nipple. In the latter direction, the MTF peak shift is more emphasized for large angular range systems (2.0 versus 1.0lp/mm). In-depth resolution of the planes, via the full width at half maximum (FWHM) from the point spread function of a 25μm tungsten wire, is not only influenced by angular range and yields 1.3-4.6mm among systems. The artifact spread function from 1mm diameter tungsten beads depends mainly on angular range, yielding two tendencies whether large (FWHM is 4.5mm) or small (FWHM is 10mm) angular range is used. DBT delivers per scan a mean glandular dose between 1.4 and 2.7mGy for a 45mm thick polymethyl methacrylate (PMMA) block. In conclusion, we have identified and analysed specific metrics that can be used for quality assurance of DBT systems.

摘要

这项工作的目的是研究和评估三种不同的数字乳腺断层合成(DBT)系统在临床环境中的性能(西门子Mammomat Inspiration、Hologic Selenia Dimensions和富士胶片Amulet Innovality)。特性描述包括对探测器、自动曝光控制以及DBT投影和重建平面分辨率的研究。使用1mm厚的钢边缘测量DBT投影的调制传递函数(MTF),显示出很强的各向异性(在管行进方向上MTF0.5频率低30 - 40%)。用25μm钨丝测量的平面内MTF0.5,在管行进方向上范围为1.3至1.8lp/mm,在胸壁 - 乳头方向上为2.4至3.7lp/mm。在后一个方向上,对于大角度范围系统,MTF峰值偏移更为明显(2.0对1.0lp/mm)。通过25μm钨丝点扩散函数的半高宽(FWHM)测量平面的深度分辨率,不仅受角度范围影响,且在各系统间为1.3 - 4.6mm。直径1mm钨珠的伪影扩散函数主要取决于角度范围,使用大角度范围(FWHM为4.5mm)或小角度范围(FWHM为10mm)会产生两种趋势。对于45mm厚的聚甲基丙烯酸甲酯(PMMA)块,DBT每次扫描的平均腺体剂量在1.4至2.7mGy之间。总之,我们已经识别并分析了可用于DBT系统质量保证的特定指标。