Dirks Niek F, Vesper Hubert W, van Herwaarden Antonius E, van den Ouweland Jody M W, Kema Ido P, Krabbe Johannes G, Heijboer Annemieke C
Department of Clinical Chemistry, Endocrine Laboratory, VU University Medical Center, Amsterdam, The Netherlands.
Centers for Disease Control and Prevention (CDC), Division of Laboratory Sciences, Atlanta, GA, United States.
Clin Chim Acta. 2016 Nov 1;462:49-54. doi: 10.1016/j.cca.2016.08.016. Epub 2016 Aug 25.
The variety of LC-MS/MS methods measuring total 25(OH)D used today is vast and the comparability among these methods is still not well assessed.
Here, we performed a comparison in samples of healthy donors between the currently routinely used 25(OH)D LC-MS/MS methods in the Netherlands and the Ghent University reference measurement procedure to address this issue (n=40). Additionally, an interlaboratory comparison in patient serum samples assessed agreement between the Dutch diagnostic methods (n=37).
The overall correlation of the routine methods for 25(OH)D with the reference measurement procedures and with the mean of all diagnostic methods was excellent (r>0.993 and r>0.989, respectively). Three out of five methods aligned perfectly with both the reference measurement procedure and the median of all methods. One of the routine methods showed a small positive bias, while another showed a small negative bias consistently in both comparisons.
The biases most probably originated from differences in calibration procedure and may be obviated by reassessing calibration of stock standards and/or calibrator matrices. In conclusion, five diagnostic centers have performed a comparison with the 25(OH)D Ghent University reference measurement procedure in healthy donor serum samples and a comparison among themselves in patient serum samples. Both analyses showed a high correlation and specificity of the routine LC-MS/MS methods, yet did reveal some small standardization issues that could not be traced back to the technical details of the different methods. Hence, this study indicates various calibration procedures can result in perfect alignment.
目前用于测量总25(OH)D的液相色谱-串联质谱(LC-MS/MS)方法种类繁多,而这些方法之间的可比性仍未得到充分评估。
在此,我们对荷兰目前常规使用的25(OH)D LC-MS/MS方法与根特大学参考测量程序在健康供体样本中进行了比较,以解决这一问题(n = 40)。此外,在患者血清样本中进行了实验室间比较,以评估荷兰诊断方法之间的一致性(n = 37)。
25(OH)D常规方法与参考测量程序以及所有诊断方法的平均值之间的总体相关性极佳(分别为r>0.993和r>0.989)。五种方法中有三种与参考测量程序和所有方法的中位数完全一致。其中一种常规方法显示出小的正偏差,而另一种在两次比较中均始终显示出小的负偏差。
这些偏差很可能源于校准程序的差异,可通过重新评估储备标准品和/或校准物基质的校准来消除。总之,五个诊断中心在健康供体血清样本中与根特大学25(OH)D参考测量程序进行了比较,并在患者血清样本中相互进行了比较。两项分析均显示常规LC-MS/MS方法具有高度相关性和特异性,但确实揭示了一些无法追溯到不同方法技术细节的小标准化问题。因此,本研究表明各种校准程序可能导致完美匹配。