Mans Anton, Schuring Danny, Arends Mark P, Vugts Cornelia A J M, Wolthaus Jochem W H, Lotz Heidi T, Admiraal Marjan, Louwe Rob J W, Öllers Michel C, van de Kamer Jeroen B
Department of Radiation Oncology, Netherlands Cancer Institute, Amsterdam, The Netherlands.
Phys Med Biol. 2016 Oct 7;61(19):7221-7235. doi: 10.1088/0031-9155/61/19/7221. Epub 2016 Sep 20.
In 2010, the NCS (Netherlands Commission on Radiation Dosimetry) installed a subcommittee to develop guidelines for quality assurance and control for volumetric modulated arc therapy (VMAT) treatments. The report (published in 2015) has been written by Dutch medical physicists and has therefore, inevitably, a Dutch focus. This paper is a condensed version of these guidelines, the full report in English is freely available from the NCS website www.radiationdosimetry.org. After describing the transition from IMRT to VMAT, the paper addresses machine quality assurance (QA) and treatment planning system (TPS) commissioning for VMAT. The final section discusses patient specific QA issues such as the use of class solutions, measurement devices and dose evaluation methods.
2010年,荷兰辐射剂量测定委员会(NCS)设立了一个小组委员会,以制定容积调强弧形治疗(VMAT)治疗的质量保证和控制指南。该报告(于2015年发布)由荷兰医学物理学家撰写,因此不可避免地以荷兰为重点。本文是这些指南的精简版,完整的英文报告可从NCS网站www.radiationdosimetry.org免费获取。在描述了从调强放疗(IMRT)到VMAT的转变之后,本文讨论了VMAT的机器质量保证(QA)和治疗计划系统(TPS)调试。最后一部分讨论了针对患者的QA问题,如类解决方案的使用、测量设备和剂量评估方法。