Evaluation of Corneal Deformation Parameters Provided by the Corvis ST Tonometer After Trabeculectomy.

作者信息

Zong Yuan, Wu Na, Fu Zhixuan, Kong Xiangmei

机构信息

*Department of Ophthalmology and Visual Science, Eye, Ear, Nose and Throat Hospital, Shanghai Medical College of Fudan University †Key Laboratory of Myopia, Ministry of Health ‡Shanghai Key Laboratory of Visual Impairment and Restoration, Fudan University, Shanghai, China.

出版信息

J Glaucoma. 2017 Feb;26(2):166-172. doi: 10.1097/IJG.0000000000000590.

Abstract

PURPOSE

The aims of this study were to evaluate how the corneal deformation parameters provided by the Corvis ST tonometer (CST) were influenced by pressure-lowering ocular surgery, and to determine the correlations of intraocular pressure (IOP) and axial length (AL) with CST corneal deformation parameters.

METHODS

This prospective 1-month study enrolled 22 subjects (22 eyes) who underwent trabeculectomy combined with mitomycin C. The corneal deformation parameters were measured using the CST. IOP was measured before and after surgery by a Goldmann applanation tonometer and the CST. The central corneal thickness and AL were also recorded. The correlations of the corneal deformation parameters with central corneal thickness, AL, and IOP changes were determined by linear regression analysis.

RESULTS

IOP decreased significantly after surgery. AL was significantly shorter at 1 week after surgery. There were significant changes in the CST parameters time 1, velocity 1, velocity 2, peak distance highest concavity, and deformation amplitude highest concavity at 1 week and 1 month after surgery and in time highest concavity at 1 week after surgery. The change in time 1 was significantly correlated with preoperative IOP and the IOP reductions, and was positively correlated with the decrease in AL at 1 week. The increase in velocity 1 was negatively correlated with preoperative IOP, and IOP reductions. The change in time highest concavity was negatively correlated with IOP before surgery and the decrease in IOP at 1 week.

CONCLUSION

CST is a good choice for measuring IOP, especially when aiming for normalization of IOP after glaucoma surgery.

摘要

目的

本研究旨在评估Corvis ST眼压计(CST)提供的角膜变形参数如何受到降低眼压的眼科手术的影响,并确定眼压(IOP)和眼轴长度(AL)与CST角膜变形参数之间的相关性。

方法

这项为期1个月的前瞻性研究纳入了22名接受小梁切除术联合丝裂霉素C治疗的受试者(22只眼)。使用CST测量角膜变形参数。术前和术后通过Goldmann压平眼压计和CST测量IOP。还记录了中央角膜厚度和AL。通过线性回归分析确定角膜变形参数与中央角膜厚度、AL和IOP变化之间的相关性。

结果

术后IOP显著降低。术后1周时AL明显缩短。术后1周和1个月时,CST参数时间1、速度1、速度2、最高凹陷处的峰值距离和最高凹陷处的变形幅度以及术后1周时最高凹陷处的时间有显著变化。时间1的变化与术前IOP和IOP降低显著相关,并且与术后1周时AL的减少呈正相关。速度1的增加与术前IOP和IOP降低呈负相关。最高凹陷处时间的变化与术前IOP和术后1周时IOP的降低呈负相关。

结论

CST是测量IOP的良好选择,尤其是在青光眼手术后旨在使IOP正常化时。