Using Radar Plots to Demonstrate the Accuracy and Precision of 6 Blood Glucose Monitoring Systems.

作者信息

Pardo Scott, Dunne Nancy, Simmons David A

机构信息

1 Ascensia Diabetes Care, Parsippany, NJ.

出版信息

J Diabetes Sci Technol. 2017 Sep;11(5):966-969. doi: 10.1177/1932296817713026. Epub 2017 Jun 12.

Abstract

BACKGROUND

Previously, fingertip capillary blood glucose measurements from the CONTOURNEXT (CN) blood glucose monitoring system (BGMS) and 5 other BGMSs were evaluated in comparison with measurements from a reference YSI glucose analyzer. Here, we use Radar Plots to graphically represent the accuracy and precision results from the previous study, including whether they met ISO 15197:2013 accuracy criteria.

METHOD

A Radar Plot, a new method for capturing a distinct, single visualization of BGMS analytical performance, is a collection of concentric circles, each representing a particular magnitude of error. The center of the plot represents zero error (BGMS result is equivalent to reference result); as points are more distant from the center, the error increases, expressed in units of mg/dL or percentage for YSI values <100 and ≥100 mg/dL, respectively. The position of the data point above or below the horizontal line bisecting the plot indicates whether the BGMS measurement error was positive (BGMS result > YSI result) or negative (BGMS result < YSI result). Points within the "15-15 Zone," representing ±15 mg/dL or ±15% error, satisfy ISO 15197:2013 accuracy criteria.

RESULTS

The percentage of results within the 15-15 Zone ranged from 83.6% to 99.8% for the 6 BGMSs (99.6% for CN).

CONCLUSIONS

Radar Plots provide a different method for visually comparing the analytical performance of multiple BGMSs. The tight clustering of data points at the center of the CN Radar Plot illustrates the analytical performance of CN compared with 5 other BGMSs.

摘要

背景

此前,对Contour Next(CN)血糖监测系统(BGMS)和其他5种BGMS进行了指尖毛细血管血糖测量,并与参考YSI葡萄糖分析仪的测量结果进行了比较。在此,我们使用雷达图以图形方式展示先前研究的准确性和精密度结果,包括它们是否符合ISO 15197:2013准确性标准。

方法

雷达图是一种用于直观呈现BGMS分析性能的新方法,它由一组同心圆组成,每个圆代表特定的误差幅度。图的中心表示零误差(BGMS结果等同于参考结果);随着点离中心越远,误差越大,对于YSI值<100和≥100 mg/dL,误差分别以mg/dL或百分比为单位表示。数据点在平分该图的水平线之上或之下的位置表明BGMS测量误差是正(BGMS结果>YSI结果)还是负(BGMS结果<YSI结果)。位于“15 - 15区域”内的点,代表±15 mg/dL或±15%的误差,符合ISO 15197:2013准确性标准。

结果

6种BGMS在15 - 15区域内的结果百分比范围为83.6%至99.8%(CN为99.6%)。

结论

雷达图提供了一种直观比较多种BGMS分析性能的不同方法。CN雷达图中心的数据点紧密聚集,说明了CN与其他5种BGMS相比的分析性能。

相似文献

6
Accuracy and user performance evaluation of a blood glucose monitoring system.血糖仪准确性和用户性能评估。
Diabetes Technol Ther. 2015 Mar;17(3):152-8. doi: 10.1089/dia.2014.0129. Epub 2014 Dec 10.
10

引用本文的文献

本文引用的文献

3
The surveillance error grid.监测误差网格。
J Diabetes Sci Technol. 2014 Jul;8(4):658-72. doi: 10.1177/1932296814539589. Epub 2014 Jun 13.

文献AI研究员

20分钟写一篇综述,助力文献阅读效率提升50倍。

立即体验

用中文搜PubMed

大模型驱动的PubMed中文搜索引擎

马上搜索

文档翻译

学术文献翻译模型,支持多种主流文档格式。

立即体验