The precision and accuracy of a quantitative magnetic resonance (EchoMRI Infants) system in newborns were determined.
Canola oil and drinking water phantoms (increments of 10 g to 1.9 kg) were scanned four times. Instrument reproducibility was assessed from three scans (within 10 minutes) in 42 healthy term newborns (12-70 hours post birth). Instrument precision was determined from the coefficient of variation (CV) of repeated scans for total water, lean mass, and fat measures for newborns and the mean difference between weight and measurement for phantoms. In newborns, the system accuracy for total body water (TBW) was tested against deuterium dilution (D O).
In phantoms, the repeatability and accuracy of fat and water measurements increased as the weight of oil and water increased. TBW was overestimated in amounts >200 g. In newborns weighing 3.14 kg, fat, lean mass, and TBW were 0.52 kg (16.48%), 2.28 kg, and 2.40 kg, respectively. EchoMRI's reproducibility (CV) was 3.27%, 1.83%, and 1.34% for total body fat, lean mass, and TBW, respectively. EchoMRI-TBW values did not differ from D O; mean difference, -1.95 ± 6.76%, P = 0.387; mean bias (limits of agreement), 0.046 kg (-0.30 to 0.39 kg).
The EchoMRI Infants system's precision and accuracy for total body fat and lean mass are better than established techniques and equivalent to D O for TBW in phantoms and newborns.
确定定量磁共振(EchoMRI婴儿型)系统在新生儿中的精密度和准确性。
对菜籽油和饮用水体模(重量从10克递增至1.9千克)进行4次扫描。在42名健康足月儿(出生后12至70小时)中,通过3次扫描(10分钟内)评估仪器的可重复性。通过新生儿重复扫描总水量、瘦体重和脂肪量的变异系数(CV)以及体模重量与测量值之间的平均差异来确定仪器的精密度。在新生儿中,采用氘稀释法(D₂O)检验全身水(TBW)的系统准确性。
在体模中,随着油和水重量的增加,脂肪和水测量的重复性和准确性提高。当重量超过200克时,TBW被高估。对于体重3.14千克的新生儿,脂肪、瘦体重和TBW分别为0.52千克(16.48%)、2.28千克和2.40千克。EchoMRI对全身脂肪、瘦体重和TBW的可重复性(CV)分别为3.27%、1.83%和1.34%。EchoMRI-TBW值与D₂O值无差异;平均差异为-1.95±6.76%,P = 0.387;平均偏差(一致性界限)为0.046千克(-0.30至0.39千克)。
EchoMRI婴儿型系统对全身脂肪和瘦体重的精密度和准确性优于既定技术,在体模和新生儿中对TBW的测量与氘稀释法相当。