Grover-Páez Fernando, Cardona-Muñoz Ernesto G, Cardona-Müller David, Guzmán-Saldívar Víctor H, Rodríguez-De la Cerda Mariana, Jiménez-Cázarez Mayra B, Totsuka-Sutto Sylvia E, Alanis-Sánchez Guillermo A, Ramos-Becerra Carlos G
Arterial Stiffness Laboratory, Department of Physiology, University of Guadalajara, Guadalajara, Mexico.
Blood Press Monit. 2017 Dec;22(6):375-378. doi: 10.1097/MBP.0000000000000290.
The aim of this study was to determine the accuracy of the Omron HEM-7320-LA with Intelli Wrap technology cuff HEM-FL1 for self-measurement and clinic blood pressure (BP) measurement according to the European Society of Hypertension International Protocol revision 2010.
The evaluation was performed in 39 individuals. The mean age of the participants was 47.9±14 years; systolic BP was 145.2±24.3 mmHg (range: 97-190), diastolic BP was 90.9±12.9 mmHg (range: 68-120), and arm circumference was 30.8±4 cm (range: 25-38.5).
The device successfully fulfilled the established criteria of the validation protocol. The device overestimated systolic BP by 0.6±5.7 mmHg and diastolic BP by 2.2±5.1 mmHg. The specially designed cuff HEM-FL1 to cover a broad range of arm circumferences and self-placement fulfilled the requirements of the International Protocol.