Clinicians have been slow to adopt digital impression technologies due possibly to perceived technique sensitivities involved in data acquisition. This research has two aims: determine whether scan pattern and sequence affects the accuracy of the three-dimensional (3D) model created from this digital impression and to compare the 5 imaging systems with regards to their scanning accuracy for sextant impressions.
Six digital intraoral impression systems were used to scan a typodont sextant with optical properties similar to natural teeth. The impressions were taken using five different scan patterns and the resulting digital models were overlayed on a master digital model to determine the accuracy of each scanner performing each scan pattern. Furthermore, regardless of scan pattern, each digital impression system was evaluated for accuracy to the other systems in this same manner.
No differences of significance were noted in the accuracy of 3D models created using six distinct scan patterns with one exception involving the CEREC Omnicam. Planmeca Planscan was determined to be the truest scanner while 3Shape Trios was determined to be the most precise for sextant impression making.
Scan pattern does not significantly affect the accuracy of the resulting digital model for sextant scanning.
Companies who make digital impression systems often recommend a scan pattern specific for their system. However, every clinical scanning scenario is different and may require a different approach. Knowing how important scan pattern is with regards to accuracy would be helpful for guiding a growing number of practitioners who are utilizing this technology.
由于可能涉及到数据采集方面的技术敏感性,临床医生在采用数字化印模技术方面一直较为缓慢。本研究有两个目的:确定扫描模式和顺序是否会影响从数字印模创建的三维(3D)模型的准确性,并比较这 5 种成像系统在 sextant 印模方面的扫描准确性。
使用 6 种数字化口内印模系统扫描与天然牙齿光学性质相似的典型 sextant。采用 5 种不同的扫描模式进行印模扫描,将所得数字模型与主数字模型叠加,以确定每个扫描仪执行每种扫描模式的准确性。此外,无论扫描模式如何,均以相同方式评估每个数字印模系统与其他系统的准确性。
在使用六种不同的扫描模式创建的 3D 模型的准确性方面,没有显著差异,只有涉及 CEREC Omnicam 的情况除外。Planmeca Planscan 被确定为最准确的扫描仪,而 3Shape Trios 被确定为 sextant 印模制作最精确的扫描仪。
扫描模式不会显著影响 sextant 扫描的数字模型的准确性。
制造数字化印模系统的公司通常会为其系统推荐特定的扫描模式。然而,每个临床扫描场景都是不同的,可能需要不同的方法。了解扫描模式在准确性方面的重要性将有助于指导越来越多使用这项技术的从业者。