在近视儿童中,与 Pentacam AXL 和 IOLMaster 700 相比,使用 MYAH 设备测量的眼轴长度和眼前节参数的一致性。

Agreement of axial length and anterior segment parameters measured with the MYAH device compared to Pentacam AXL and IOLMaster 700 in myopic children.

机构信息

Department of Ophthalmology, University of Health Sciences, Dıskapı Yıldırım Beyazıt Training and Research Hospital, Ankara, Turkey.

Kaskaloglu Eye Hospital, Izmir, Turkey.

出版信息

Int Ophthalmol. 2023 Feb;43(2):475-482. doi: 10.1007/s10792-022-02444-w. Epub 2022 Jul 31.

Abstract

PURPOSE

To compare the difference and agreement of axial length (AL) and anterior segment parameters obtained from the MYAH device with Pentacam AXL and IOLMaster 700 in myopic children.

METHODS

The study included 60 eyes of 60 myopic children. AL, keratometry (K), and horizontal corneal diameter (CD) were measured with Pentacam AXL, IOLMaster 700, and MYAH, respectively. The Friedman test was used to assess the differences. The Intraclass Correlation Coefficient (ICC) and Bland-Altman plots were used to assess the consistency of measurements.

RESULTS

The mean age was 10.2 ± 1.8 years (7-16 years). No statistically significant difference was determined between the Pentacam AXL, IOLMaster 700, and MYAH devices in terms of mean AL values (23.61 ± 1.42, 23.62 ± 1.45, 23.61 ± 1.42, respectively) (p = 0.06). The difference between devices in the mean steep K, flat K, mean K, and CD was statistically significant but clinically insignificant (steep K; 44.45 ± 1.25, 44.59 ± 1.23, 44.51 ± 1.24, flat K; 43.29 ± 1.28, 43.43 ± 1.29, 43.35 ± 1.30, mean K; 43.85 ± 1.21, 44.00 ± 1.19, 43.94 ± 1.20, and CD; 11.90 ± 0.34, 12.11 ± 0.38, 11.96 ± 0.31, respectively; p < 0.05). ICC and Bland-Altman plot analysis revealed a high correlation between the three devices in AL, steep K, flat K, mean K, and CD measurements.

CONCLUSION

There was a quite good agreement between the MYAH, Pentacam AXL, and IOLMaster 700 devices regarding AL and anterior segment parameters. MYAH provides reliable measurements and will be a good option in the diagnosis of and follow-up with myopic children.

摘要

目的

比较 MYAH 设备与 Pentacam AXL 和 IOLMaster 700 在近视儿童中测量的眼轴(AL)和眼前节参数的差异和一致性。

方法

本研究纳入 60 例(60 只眼)近视儿童。分别用 Pentacam AXL、IOLMaster 700 和 MYAH 测量 AL、角膜曲率(K)和水平角膜直径(CD)。采用 Friedman 检验评估差异,采用组内相关系数(ICC)和 Bland-Altman 图评估测量值的一致性。

结果

平均年龄为 10.2±1.8 岁(7-16 岁)。Pentacam AXL、IOLMaster 700 和 MYAH 设备在平均 AL 值(分别为 23.61±1.42、23.62±1.45、23.61±1.42)方面无统计学差异(p=0.06)。设备之间在平均陡峭 K、平坦 K、平均 K 和 CD 方面的差异具有统计学意义,但临床意义不大(陡峭 K:44.45±1.25、44.59±1.23、44.51±1.24,平坦 K:43.29±1.28、43.43±1.29、43.35±1.30,平均 K:43.85±1.21、44.00±1.19、43.94±1.20,CD:11.90±0.34、12.11±0.38、11.96±0.31;p<0.05)。ICC 和 Bland-Altman 图分析显示,这三种设备在 AL、陡峭 K、平坦 K、平均 K 和 CD 测量值之间具有高度相关性。

结论

在 AL 和眼前节参数方面,MYAH 与 Pentacam AXL 和 IOLMaster 700 设备之间具有相当好的一致性。MYAH 提供可靠的测量值,是诊断和随访近视儿童的良好选择。

https://cdn.ncbi.nlm.nih.gov/pmc/blobs/ada4/9340674/c150a1abba84/10792_2022_2444_Fig1_HTML.jpg

相似文献

2
Repeatability and agreement of Scheimpflug-based and swept-source optical biometry measurements.
Cont Lens Anterior Eye. 2017 Oct;40(5):318-322. doi: 10.1016/j.clae.2017.03.007. Epub 2017 Mar 23.
3
Comparison of Anterior Segment Biometric Measurements between Pentacam HR and IOLMaster in Normal and High Myopic Eyes.
PLoS One. 2015 Nov 17;10(11):e0143110. doi: 10.1371/journal.pone.0143110. eCollection 2015.
4
A Comparison of Optical Biometers Used in Children for Myopia Control.
Klin Monbl Augenheilkd. 2023 Nov;240(11):1306-1313. doi: 10.1055/a-2117-9335. Epub 2023 Jun 26.
5
[Comparison of biometry with the Pentacam AXL, IOLMaster 700 and IOLMaster 500 in cataract patients].
Zhonghua Yan Ke Za Zhi. 2019 Jul 11;55(7):515-521. doi: 10.3760/cma.j.issn.0412-4081.2019.07.007.
6
Agreement between 2 swept-source OCT biometers and a Scheimpflug partial coherence interferometer.
J Cataract Refract Surg. 2021 Apr 1;47(4):488-495. doi: 10.1097/j.jcrs.0000000000000483.
7
Effectiveness and agreement of 3 optical biometers in measuring axial length in the eyes of patients with mature cataracts.
J Cataract Refract Surg. 2020 Sep;46(9):1222-1228. doi: 10.1097/j.jcrs.0000000000000237.
8
Agreement between IOLMaster 700 and Pentacam AXL for IOL power measurement in patients with high myopia.
Indian J Ophthalmol. 2024 Jul 1;72(7):1021-1025. doi: 10.4103/IJO.IJO_1350_23. Epub 2024 Mar 8.
9
Suitability of multifunction devices Myah and Myopia Master for monitoring myopia progression in children and adults.
Ophthalmic Physiol Opt. 2024 Jul;44(5):1017-1030. doi: 10.1111/opo.13332. Epub 2024 May 17.

引用本文的文献

1
Agreement between IOLMaster 700 and Pentacam AXL for IOL power measurement in patients with high myopia.
Indian J Ophthalmol. 2024 Jul 1;72(7):1021-1025. doi: 10.4103/IJO.IJO_1350_23. Epub 2024 Mar 8.
3
Agreements' profile of Scheimpflug-based optical biometer with gold standard partial coherence interferometry.
Int J Ophthalmol. 2023 Dec 18;16(12):2095-2104. doi: 10.18240/ijo.2023.12.24. eCollection 2023.
4
A Comparison of Optical Biometers Used in Children for Myopia Control.
Klin Monbl Augenheilkd. 2023 Nov;240(11):1306-1313. doi: 10.1055/a-2117-9335. Epub 2023 Jun 26.
5
Clinical Validation of a New Optical Biometer for Myopia Control in a Healthy Pediatric Population.
Children (Basel). 2022 Nov 9;9(11):1713. doi: 10.3390/children9111713.

本文引用的文献

1
Comparison of IOL-Master 700 and IOL-Master 500 biometers in ocular biological parameters of adolescents.
Int J Ophthalmol. 2021 Jul 18;14(7):1013-1017. doi: 10.18240/ijo.2021.07.08. eCollection 2021.
2
Evaluation of 6 biometers based on different optical technologies.
J Cataract Refract Surg. 2022 Jan 1;48(1):16-25. doi: 10.1097/j.jcrs.0000000000000690.
3
Recent Epidemiology Study Data of Myopia.
J Ophthalmol. 2020 Nov 4;2020:4395278. doi: 10.1155/2020/4395278. eCollection 2020.
4
Agreement between 2 swept-source OCT biometers and a Scheimpflug partial coherence interferometer.
J Cataract Refract Surg. 2021 Apr 1;47(4):488-495. doi: 10.1097/j.jcrs.0000000000000483.
6
Effectiveness and agreement of 3 optical biometers in measuring axial length in the eyes of patients with mature cataracts.
J Cataract Refract Surg. 2020 Sep;46(9):1222-1228. doi: 10.1097/j.jcrs.0000000000000237.
10
Efficiency and measurements agreement between swept-source OCT and low-coherence interferometry biometry systems.
Graefes Arch Clin Exp Ophthalmol. 2018 Mar;256(3):559-566. doi: 10.1007/s00417-018-3909-9. Epub 2018 Feb 1.

文献AI研究员

20分钟写一篇综述,助力文献阅读效率提升50倍。

立即体验

用中文搜PubMed

大模型驱动的PubMed中文搜索引擎

马上搜索

文档翻译

学术文献翻译模型,支持多种主流文档格式。

立即体验