A convenient method to verify the accuracy of oscillometric blood pressure monitors by the auscultatory method: A smartphone-based app.

机构信息

Department of Cardiology, Shanghai General Hospital, Shanghai Jiaotong University School of Medicine, Shanghai, China.

Department of Cardiology, Huai'an First People's Hospital affiliated with Nanjing Medical University, Huai'an, China.

出版信息

J Clin Hypertens (Greenwich). 2019 Feb;21(2):173-180. doi: 10.1111/jch.13460. Epub 2018 Dec 20.

Abstract

It is recommended that oscillometric devices be calibrated by auscultation when first used, but this is difficult in practice. Here, we introduce a smartphone-based technique to verify the accuracy of blood pressure monitors (BPMs). We enrolled 99 consecutive subjects and tested 6 brands of BPMs in this study. During measurements of electronic oscillometric BPMs, Korotkoff sounds were simultaneously collected using a stethoscope head beneath a cuff connected to a smartphone, and an app named Accutension Stetho could then yield an auscultatory BP reading as a reference. Next, differences in BP between the different BPMs and Accutension Stetho were determined. The percentage of BP differences falling within 5, 10, and 15 mm Hg; the mean (MD) value; and the standard deviation (SD) of BP differences and deflation errors were analyzed among all the BPMs. We found that the percentages of SBP differences falling within 5 mm Hg of the 6 BPMs were 80%, 79%, 77%, 72%, 68%, and 63%, in turn. The deflation rates among the 6 BPMs were 2.23, 3.48, 6.10, 2.44, 3.66, and 4.85 mm Hg/beat, respectively. Deflation errors, which were defined as deflation prior to the end of the Korotkoff sounds, existed in 4 BPMs. In conclusion, Accutension Stetho could detect BP differences between oscillometric BPM readings and simultaneous auscultatory readings. Diastolic BP was overestimated when the device deflated prior to the end of the Korotkoff sounds. Using the app, it is possible to evaluate the accuracy of BPMs among the same subjects.

摘要

建议在首次使用时通过听诊法对示波法设备进行校准,但实际操作中这很困难。在这里,我们介绍了一种基于智能手机的技术,用于验证血压计(BPM)的准确性。我们在这项研究中纳入了 99 名连续受试者并测试了 6 个品牌的 BPM。在电子示波法 BPM 的测量过程中,同时使用听诊器头在与智能手机相连的袖带下收集柯氏音,然后名为 Accutension Stetho 的应用程序可以产生听诊血压读数作为参考。接下来,确定不同 BPM 与 Accutension Stetho 之间的血压差异。分析所有 BPM 的血压差异百分比、MD 值、血压差异和放气误差的标准差。我们发现,6 个 BPM 的收缩压差异在 5mmHg 以内的百分比分别为 80%、79%、77%、72%、68%和 63%。6 个 BPM 的放气率分别为 2.23、3.48、6.10、2.44、3.66 和 4.85mmHg/beat。在 4 个 BPM 中存在在柯氏音结束前就开始放气的放气误差。总之,Accutension Stetho 可以检测示波法 BPM 读数与同时听诊读数之间的血压差异。当设备在柯氏音结束前开始放气时,舒张压会被高估。使用该应用程序,可以在同一受试者中评估 BPM 的准确性。

相似文献

本文引用的文献

5
Patient-Specific Oscillometric Blood Pressure Measurement.针对个体患者的示波法血压测量。
IEEE Trans Biomed Eng. 2016 Jun;63(6):1220-1228. doi: 10.1109/TBME.2015.2491270. Epub 2015 Oct 15.
6
Oscillometric Blood Pressure Estimation: Past, Present, and Future.血压的测振法估计:过去、现在和未来。
IEEE Rev Biomed Eng. 2015;8:44-63. doi: 10.1109/RBME.2015.2434215. Epub 2015 May 15.
7
Blood pressure estimation using maximum slope of oscillometric pulses.
Annu Int Conf IEEE Eng Med Biol Soc. 2012;2012:3239-42. doi: 10.1109/EMBC.2012.6346655.

文献检索

告别复杂PubMed语法,用中文像聊天一样搜索,搜遍4000万医学文献。AI智能推荐,让科研检索更轻松。

立即免费搜索

文件翻译

保留排版,准确专业,支持PDF/Word/PPT等文件格式,支持 12+语言互译。

免费翻译文档

深度研究

AI帮你快速写综述,25分钟生成高质量综述,智能提取关键信息,辅助科研写作。

立即免费体验