Mini EXPLORER II:一款用于宠物全身和人脑扫描的高灵敏度 PET/CT 扫描仪原型机。

Mini EXPLORER II: a prototype high-sensitivity PET/CT scanner for companion animal whole body and human brain scanning.

机构信息

Molecular Imaging Business Unit, Shanghai United Imaging Healthcare, Co. Ltd., Shanghai, People's Republic of China.

出版信息

Phys Med Biol. 2019 Mar 21;64(7):075004. doi: 10.1088/1361-6560/aafc6c.

Abstract

As part of the EXPLORER total-body positron emission tomography (PET) project, we have designed and built a high-resolution, high-sensitivity PET/CT scanner, which is expected to have excellent performance for companion animal whole body and human brain imaging. The PET component has a ring diameter of 52 cm and an axial field of view of 48.3 cm. The detector modules are composed of arrays of lutetium (yttrium) oxyorthosilicate (LYSO) crystals of dimensions 2.76  ×  2.76  ×  18.1 mm coupled to silicon photomultipliers (SiPMs) for read-out. The CT component is a 24 detector row CT scanner with a 50 kW x-ray tube. PET system time-of-flight resolution was measured to be 409  ±  39 ps and average system energy resolution was 11.7%  ±  1.5% at 511 keV. The NEMA NU2-2012 system sensitivity was found to be 52-54 kcps MBq. Spatial resolution was 2.6 mm at 10 mm from the center of the FOV and 2.0 mm rods were clearly resolved on a mini-Derenzo phantom. Peak noise-equivalent count (NEC) rate, using the NEMA NU 2-2012 phantom, was measured to be 314 kcps at 9.2 kBq cc. The CT scanner passed the technical components of the American College of Radiology (ACR) accreditation tests. We have also performed scans of a Hoffman brain phantom and we show images from the first canine patient imaged on this device.

摘要

作为 EXPLORER 全身正电子发射断层扫描 (PET) 项目的一部分,我们设计并制造了一种高分辨率、高灵敏度的 PET/CT 扫描仪,预计在用于宠物全身和人脑成像方面具有出色的性能。PET 组件的环直径为 52cm,轴向视野为 48.3cm。探测器模块由尺寸为 2.76×2.76×18.1mm 的掺铥硅酸钇 (LYSO) 晶体阵列组成,并与硅光电倍增管 (SiPM) 耦合进行读出。CT 组件是一个 24 探测器排 CT 扫描仪,配备一个 50kW X 射线管。PET 系统飞行时间分辨率测量值为 409 ± 39 ps,在 511keV 时平均系统能量分辨率为 11.7% ± 1.5%。发现 NEMA NU2-2012 系统灵敏度为 52-54 kcps MBq。在距视场中心 10mm 处的空间分辨率为 2.6mm,在 Mini-Derenzo 体模上可以清楚地分辨出 2.0mm 棒。使用 NEMA NU 2-2012 体模测量的峰值噪声等效计数 (NEC) 率为 314 kcps,在 9.2kBq cc 时。CT 扫描仪通过了美国放射学院 (ACR) 的技术组件认证测试。我们还对 Hoffman 脑体模进行了扫描,并展示了在该设备上对第一只犬患者进行成像的图像。

https://cdn.ncbi.nlm.nih.gov/pmc/blobs/2919/6666419/34b36abda9d8/nihms-1041249-f0001.jpg

相似文献

8
Performance Characteristics of the Digital Biograph Vision PET/CT System.Digital Biograph Vision PET/CT 系统的性能特点。
J Nucl Med. 2019 Jul;60(7):1031-1036. doi: 10.2967/jnumed.118.215418. Epub 2019 Jan 10.
10
Performance evaluation of the new whole-body PET/CT scanner: Discovery ST.新型全身PET/CT扫描仪Discovery ST的性能评估
Eur J Nucl Med Mol Imaging. 2004 Jun;31(6):867-81. doi: 10.1007/s00259-003-1444-2. Epub 2004 Feb 10.

引用本文的文献

10
Hybrid total-body pet scanners-current status and future perspectives.混合式全身 PET 扫描仪:现状与未来展望。
Eur J Nucl Med Mol Imaging. 2022 Jan;49(2):445-459. doi: 10.1007/s00259-021-05536-4. Epub 2021 Oct 14.

本文引用的文献

7
NEUROSCIENCE. It takes the world to understand the brain.神经科学。需要全世界共同努力来理解大脑。
Science. 2015 Oct 2;350(6256):42-4. doi: 10.1126/science.aad4120. Epub 2015 Oct 1.
8
Research priorities. The NIH BRAIN Initiative.研究重点。NIH BRAIN 计划。
Science. 2013 May 10;340(6133):687-8. doi: 10.1126/science.1239276.
9
Technical performance evaluation of a human brain PET/MRI system.人脑 PET/MRI 系统的技术性能评估。
Eur Radiol. 2012 Aug;22(8):1776-88. doi: 10.1007/s00330-012-2415-4. Epub 2012 Mar 27.

文献检索

告别复杂PubMed语法,用中文像聊天一样搜索,搜遍4000万医学文献。AI智能推荐,让科研检索更轻松。

立即免费搜索

文件翻译

保留排版,准确专业,支持PDF/Word/PPT等文件格式,支持 12+语言互译。

免费翻译文档

深度研究

AI帮你快速写综述,25分钟生成高质量综述,智能提取关键信息,辅助科研写作。

立即免费体验