Quantitative lung ultrasonography: a putative new algorithm for automatic detection and quantification of B-lines.

机构信息

Anaesthesia and Intensive Care Unit, E.O. Ospedali Galliera, Genova, Italy.

Department of Surgical Sciences and Integrated Diagnostics, University of Genoa, Genoa, Italy.

出版信息

Crit Care. 2019 Aug 28;23(1):288. doi: 10.1186/s13054-019-2569-4.

Abstract

BACKGROUND

This pilot study was designed to develop a fully automatic and quantitative scoring system of B-lines (QLUSS: quantitative lung ultrasound score) involving the pleural line and to compare it with previously described semi-quantitative scores in the measurement of extravascular lung water as determined by standard thermo-dilution.

METHODS

This was a prospective observational study of 12 patients admitted in the intensive care unit with acute respiratory distress and each provided with 12 lung ultrasound (LUS) frames. Data collected from each patient consisted in five different scores, four semi-quantitative (nLUSS, cLUSS, qLUSS, %LUSS) and quantitative scores (QLUSS). The association between LUS scores and extravascular lung water (EVLW) was determined by simple linear regression (SLR) and robust linear regression (RLR) methods. A correlation analysis between the LUS scores was performed by using the Spearman rank test. Inter-observer variability was tested by computing intraclass correlation coefficient (ICC) in two-way models for agreement, basing on scores obtained by different raters blinded to patients' conditions and clinical history.

RESULTS

In the SLR, QLUSS showed a stronger association with EVLW (R = 0.57) than cLUSS (R = 0.45) and nLUSS (R = 0.000), while a lower association than qLUSS (R = 0.85) and %LUSS (R = 0.72) occurred. By applying RLR, QLUSS showed an association for EVLW (R = 0.86) comparable to qLUSS (R = 0.85) and stronger than %LUSS (R = 0.72). QLUSS was significantly correlated with qLUSS (r = 0.772; p = 0.003) and %LUSS (r = 0.757; p = 0.005), but not with cLUSS (r = 0.561; p = 0.058) and nLUSS (r = 0.105; p = 0.744). Moreover, QLUSS showed the highest ICC (0.998; 95%CI from 0.996 to 0.999) among the LUS scores.

CONCLUSIONS

This study demonstrates that computer-aided scoring of the pleural line percentage affected by B-lines has the potential to assess EVLW. QLUSS may have a significant impact, once validated with a larger dataset composed by multiple real-time frames. This approach has the potentials to be advantageous in terms of faster data analysis and applicability to large sets of data without increased costs. On the contrary, it is not useful in pleural effusion or consolidations.

摘要

背景

本初步研究旨在开发一种完全自动和定量的 B 线评分系统(QLUSS:定量肺超声评分),涉及胸膜线,并将其与以前描述的半定量评分进行比较,以测量标准热稀释法确定的血管外肺水。

方法

这是一项对 12 名入住重症监护病房的急性呼吸窘迫患者进行的前瞻性观察性研究,每位患者提供 12 个肺超声(LUS)帧。从每位患者收集的数据包括五个不同的评分,四个半定量(nLUSS、cLUSS、qLUSS、%LUSS)和定量评分(QLUSS)。通过简单线性回归(SLR)和稳健线性回归(RLR)方法确定 LUS 评分与血管外肺水(EVLW)之间的关系。使用 Spearman 秩检验对 LUS 评分进行相关性分析。通过计算不同评分者之间的组内相关系数(ICC),在双向模型中测试观察者间的变异性,评分者对患者情况和临床病史不知情。

结果

在 SLR 中,QLUSS 与 EVLW 的相关性(R=0.57)强于 cLUSS(R=0.45)和 nLUSS(R=0.000),而弱于 qLUSS(R=0.85)和 %LUSS(R=0.72)。通过应用 RLR,QLUSS 与 EVLW 的相关性(R=0.86)与 qLUSS(R=0.85)相当,且强于 %LUSS(R=0.72)。QLUSS 与 qLUSS(r=0.772;p=0.003)和 %LUSS(r=0.757;p=0.005)显著相关,但与 cLUSS(r=0.561;p=0.058)和 nLUSS(r=0.105;p=0.744)不相关。此外,QLUSS 在 LUS 评分中显示出最高的 ICC(0.998;95%CI 为 0.996 至 0.999)。

结论

本研究表明,计算机辅助胸膜线 B 线百分率评分有可能评估 EVLW。一旦用由多个实时帧组成的更大数据集进行验证,QLUSS 可能具有重要意义。这种方法在数据分析速度更快和适用于大数据集方面具有优势,而不会增加成本。相反,它在胸腔积液或实变中没有用。