We compared axial length (AL) measurements using spectral-domain optical coherence tomography (SD-OCT)-based and partial coherence interferometry (PCI)-based biometry in eyes with various chorioretinal diseases. Medical records were retrospectively analyzed in 123 eyes from 73 patients who underwent AL measurements using SD-OCT and PCI. AL values were highly consistent by both measurements (ICC = 0.996 in SD-OCT and 0.995 in PCI-based biometry). The mean difference in AL between the two instruments was - 0.09 ± 0.17 mm, which was significant (P < 0.001). There was a strong correlation between the measurements obtained using the two instruments (r = 0.982, P < 0.001). The 95% limits of agreement were - 0.43 to 0.25 mm, and the agreement of AL in the Bland-Altman analysis showed a difference. The difference in measurements was not associated with the difference in central retinal thickness (r = 0.033, P = 0.720). However, the presence of an epiretinal membrane affected the difference in measurements, unlike in other chorioretinal diseases (P = 0.014). SD-OCT-based biometry showed a good correlation with PCI-based biometry in the measurement of AL; however, the mean AL between the two instruments was significantly different. Therefore, AL measurements made using SD-OCT-based and PCI-based biometry are not interchangeable.
我们比较了在各种脉络膜视网膜疾病的眼中使用基于谱域光相干断层扫描(SD-OCT)和基于部分相干干涉测量(PCI)的生物测量的眼轴长度(AL)测量值。对 73 例患者的 123 只眼进行了回顾性分析,这些患者使用 SD-OCT 和 PCI 进行了 AL 测量。两种测量方法的 AL 值高度一致(SD-OCT 的 ICC 为 0.996,基于 PCI 的生物测量的 ICC 为 0.995)。两种仪器之间的 AL 平均差异为-0.09±0.17mm,差异显著(P<0.001)。两种仪器获得的测量值之间存在很强的相关性(r=0.982,P<0.001)。95%的一致性界限为-0.43 至 0.25mm,Bland-Altman 分析中的 AL 一致性显示存在差异。测量值的差异与视网膜中央厚度的差异无关(r=0.033,P=0.720)。然而,与其他脉络膜视网膜疾病不同,存在视网膜内膜会影响测量值的差异(P=0.014)。SD-OCT 基于生物测量与 PCI 基于生物测量在 AL 的测量中具有良好的相关性;然而,两种仪器之间的平均 AL 存在显著差异。因此,SD-OCT 基于生物测量和 PCI 基于生物测量的 AL 测量值不能互换。