Fişuş Andreea D, Hirnschall Nino D, Findl Oliver
From the Department of Ophthalmology, Vienna Institute for Research in Ocular Surgery, Karl Landsteiner Institute, Hanusch Hospital, Vienna, Austria.
J Cataract Refract Surg. 2021 Jan 1;47(1):87-92. doi: 10.1097/j.jcrs.0000000000000373.
To compare 2 swept-source optical coherence tomography (SS-OCT) biometers, IOLMaster 700 (biometer A), and ANTERION (biometer B).
Department of Ophthalmology, Vienna Institute for Research in Ocular Surgery, Karl Landsteiner Institute, Hanusch Hospital, Vienna, Austria.
Retrospective study.
Biometric measurements of cataract patients performed between June and July 2019 were reviewed. Patients scheduled for cataract surgery were measured with both SS-OCT devices on the same day. The following biometry parameters were compared: keratometry, central corneal thickness, anterior chamber depth (ACD), lens thickness (LT), and axial length (AL).
In total, 389 eyes of 209 subjects were measured with both devices. The mean absolute difference between the keratometry data of the 2 devices was 0.04 ± 0.05 mm (7.80 ± 0.26 mm for biometer A and 7.82 ± 0.26 mm for biometer B; P < .0001) for the steep keratometry readings and 0.04 ± 0.04 mm (7.63 ± 0.26 mm and 7.65 ± 0.25 mm; P < .0001) for the flat keratometry readings. For ACD and LT, the mean absolute difference was 0.07 ± 0.04 mm and 0.07 ± 0.04 mm. The mean absolute difference for AL was 0.02 ± 0.03 mm (23.55 ± 1.18 mm for biometer A and 23.54 ± 1.18 mm for biometer B; P < .0001).
Good agreement was found between the 2 devices, with a minor offset for ACD and LT measurements. Although differences were found to be small, the devices should not be used interchangeably.
比较两款扫频光学相干断层扫描(SS-OCT)生物测量仪,IOLMaster 700(生物测量仪A)和ANTERION(生物测量仪B)。
奥地利维也纳汉努施医院卡尔·兰德施泰纳研究所眼外科研究维也纳研究所眼科。
回顾性研究。
回顾2019年6月至7月对白内障患者进行的生物测量。计划进行白内障手术的患者在同一天使用两台SS-OCT设备进行测量。比较以下生物测量参数:角膜曲率、中央角膜厚度、前房深度(ACD)、晶状体厚度(LT)和眼轴长度(AL)。
总共对209名受试者的389只眼睛使用两台设备进行了测量。两台设备角膜曲率数据的平均绝对差值,对于陡峭角膜曲率读数为0.04±0.05毫米(生物测量仪A为7.8±0.26毫米,生物测量仪B为7.82±0.26毫米;P<.0001),对于平坦角膜曲率读数为0.04±0.04毫米(7.63±0.26毫米和7.65±0.25毫米;P<.0001)。对于ACD和LT,平均绝对差值为0.07±0.04毫米和0.07±0.04毫米。AL的平均绝对差值为0.02±0.03毫米(生物测量仪A为23.55±1.18毫米,生物测量仪B为23.54±1.18毫米;P<.0001)。
两台设备之间一致性良好,ACD和LT测量存在较小偏差。虽然发现差异较小,但两台设备不应互换使用。