Zhu Yuanfei, Yang Ming, Zhang Yan, Meng Fan, Yang Tianyue, Fang Zhiwei
Department of Instrument Science and Engineering, Shanghai Jiao Tong University, Shanghai, China (mainland).
Department of Cardiovascular Surgery, Cardiovascular Institute and Fu Wai Hospital, Chinese Academy of Medical Sciences and Peking Union Medical College, Beijing, China (mainland).
Med Sci Monit. 2020 Sep 17;26:e925367. doi: 10.12659/MSM.925367.
BACKGROUND Left ventricular assist devices (LVADs) with counter-pulsation mode have been widely used to support left ventricular function and improve coronary circulation. However, the frequency characteristics of the coronary system have not been considered. The aim of this study was to investigate the effects of pulsatile frequency of LVADs on coronary perfusion. MATERIAL AND METHODS First, a lumped parameter (LP) model incorporating coronary circulation, systemic circulation, left heart, and LVAD was established to simulate the cardiovascular system. Then, the frequency characteristics of the coronary system were analyzed and the calculation results showed that the pulsatile frequency of the LVAD has a substantial effect on coronary blood flow. To verify the accuracy of the theoretical analysis, the hemodynamic effects of the LVAD on the coronary artery were compared under 4 support modes: co-pulsation mode, and counter-pulsation modes in synchronization ratios of 1: 1, 2: 1, and 3: 1. RESULTS We found that the coronary flow increased by 5% when the working mode changed from co-pulsation to counter-pulsation in a synchronization ratio of 1: 1, and by an additional 6% when the working mode changed from counter-pulsation in a synchronization ratio of 1: 1 to counter-pulsation in a synchronization ratio of 3: 1. CONCLUSIONS This work provides a useful method to increase coronary perfusion and may be beneficial for improving myocardial function in patients with end-stage heart failure, especially those with ischemic cardiomyopathy (ICM).
具有反搏模式的左心室辅助装置(LVAD)已被广泛用于支持左心室功能并改善冠状动脉循环。然而,尚未考虑冠状动脉系统的频率特性。本研究的目的是探讨LVAD的搏动频率对冠状动脉灌注的影响。
首先,建立一个包含冠状动脉循环、体循环、左心和LVAD的集总参数(LP)模型来模拟心血管系统。然后,分析冠状动脉系统的频率特性,计算结果表明LVAD的搏动频率对冠状动脉血流有显著影响。为验证理论分析的准确性,在4种支持模式下比较LVAD对冠状动脉的血流动力学影响:共搏模式,以及同步比为1:1、2:1和3:1的反搏模式。
我们发现,当工作模式从共搏模式变为同步比为1:1的反搏模式时,冠状动脉血流量增加5%,当工作模式从同步比为1:1的反搏模式变为同步比为3:1的反搏模式时,冠状动脉血流量又增加6%。
本研究提供了一种增加冠状动脉灌注的有用方法,可能有助于改善终末期心力衰竭患者,尤其是缺血性心肌病(ICM)患者的心肌功能。