使用Sensititre自动读数仪进行快速最低抑菌浓度(MIC)检测。

Rapid MIC testing with the sensititre autoreader.

作者信息

Staneck J L, Allen S D, Harris E E, Tilton R C

机构信息

Department of Pathology and Laboratory Medicine, University of Cincinnati College of Medicine, Ohio 45267-0714.

出版信息

J Clin Microbiol. 1988 Jan;26(1):1-7. doi: 10.1128/jcm.26.1.1-7.1988.

Abstract

Automated microdilution MIC results, obtained with the Autoreader (Sensititre, Inc., Salem, N.H.) following 5 h of incubation, were compared with manually read, concurrent control MICs following 18 h of incubation in a three-laboratory comparative study. A total of 704 members of the family Enterobacteriaceae or similar gram-negative organisms were tested against 17 antimicrobial agents. Autoreader MICs were within 1 doubling dilution of control values in 92.9% of instances. Discrepancies of +/- 2 doubling dilutions and +/- 3 or greater doubling dilutions were noted in 4.5 and 2.6%, respectively, of the 7,687 drug-organism combinations analyzed. The majority of errors occurred when beta-lactam antimicrobial agents were tested with a variety of different species. MICs at 5 h, when Pseudomonas aeruginosa was used, were possible in only half the isolates tested and yielded data on only a limited number of drugs in the remaining instances. Excluding results obtained with penicillin and ampicillin, which were uniformly poor, Staphylococcus aureus Autoreader values were within +/- 1 doubling dilution of control values in 93.6% of instances, 5.4% varied by +/- 2 dilutions, and only 1% of test values by +/- 3 or more dilutions from control values for 82 isolates tested against nine antimicrobial agents. Of eight additional S. aureus isolates tested that were resistant to methicillin, only one was read correctly by the Autoreader, with results on the remaining seven appearing as either insufficient growth or as total resistance to all drugs tested. Interlaboratory reproducibility was excellent for selected isolates of S. aureus and gram-negative bacilli. The accuracy of the Sensititre Autoreader MIC results was comparable to that of other same-day quantitative systems for members of the family Enterobacteriaceae and S. aureus, while the economic and procedural advantages of the broth microdilution method was maintained.

摘要

在一项三实验室比较研究中,将经5小时孵育后用自动读数仪(Sensititre公司,新罕布什尔州塞勒姆)获得的自动化微量稀释法最低抑菌浓度(MIC)结果,与经18小时孵育后人工读取的同期对照MIC结果进行比较。总共对704株肠杆菌科细菌或类似革兰氏阴性菌进行了17种抗菌药物的测试。在92.9%的情况下,自动读数仪测得的MIC值与对照值相差在1个稀释倍数之内。在分析的7687种药物-菌株组合中,分别有4.5%和2.6%的情况出现了±2个稀释倍数以及±3个或更大稀释倍数的差异。大多数误差发生在使用β-内酰胺类抗菌药物对多种不同菌种进行测试时。当使用铜绿假单胞菌时,仅半数受试菌株能够在5小时时获得MIC值,其余情况下仅能得到有限几种药物的数据。排除青霉素和氨苄西林(其结果均较差)的结果后,针对9种抗菌药物测试的82株金黄色葡萄球菌中,自动读数仪测得的值在93.6%的情况下与对照值相差在±1个稀释倍数之内,5.4%相差±2个稀释倍数,只有1%的测试值与对照值相差±3个或更多稀释倍数。在另外测试的8株耐甲氧西林金黄色葡萄球菌中,自动读数仪仅正确读取了1株,其余7株的结果显示生长不足或对所有测试药物均表现为完全耐药。对于选定的金黄色葡萄球菌和革兰氏阴性杆菌菌株,实验室间的重现性极佳。对于肠杆菌科细菌和金黄色葡萄球菌,Sensititre自动读数仪MIC结果的准确性与其他当日定量系统相当,同时保持了肉汤微量稀释法的经济和操作优势。

相似文献

本文引用的文献

6
Clinical evaluation of new AMS GSC plus card.新型AMS GSC plus卡的临床评估
Diagn Microbiol Infect Dis. 1983 Jun;1(2):139-43. doi: 10.1016/0732-8893(83)90043-3.

文献AI研究员

20分钟写一篇综述,助力文献阅读效率提升50倍。

立即体验

用中文搜PubMed

大模型驱动的PubMed中文搜索引擎

马上搜索

文档翻译

学术文献翻译模型,支持多种主流文档格式。

立即体验