Acta Orthop. 2021 Jun;92(3):316-322. doi: 10.1080/17453674.2021.1879513. Epub 2021 Feb 1.
Background and purpose - The use of crosslinked polyethylene in total hip arthroplasty (THA) has decreased wear remarkably. It has been suggested that the antioxidative effects of vitamin E may enhance the wear properties of polyethylene even further. This study evaluates revision rates between vitamin E-infused polyethylene liners (E1 and E-poly, ZimmerBiomet, Warsaw, IN, USA) versus moderately crosslinked polyethylene (ModXLPE) liners from the same manufacturer used in primary THA.Patients and methods - We conducted a study based on data from the Finnish Arthroplasty Register. The study group consisted of 2,723 THAs with a vitamin E-infused liner and a reference group of 2,707 THAs with a moderately crosslinked polyethylene liner. Survivorship, revision risk, and re-revision causes were compared between groups.Results - The 7-year survival of the vitamin E-infused polyethylene liner group and of the reference group with revision for any reason as the endpoint was comparable (94% [95% CI 92.9-94.9] and 93% [CI 91.9-93.9], respectively). The adjusted hazard ratio (HR) for any revision was similar between the groups (0.7 [CI 0.4-1.1]). When revision for aseptic loosening was studied as the endpoint, the survival for the study group was 99% (CI 98.6-99.4) and for the reference group 99% (CI 98.7-99.5), and the risk of revision was comparable between the study groups (HR 1.3 [CI 0.7-2.5]).Interpretation - After an observation period of 7 years vitamin E-infused liners shows results equal to results obtained with crosslinked polyethylene liners.
背景与目的-在全髋关节置换术(THA)中使用交联聚乙烯可显著减少磨损。有人提出,维生素 E 的抗氧化作用可能会进一步提高聚乙烯的耐磨性能。本研究评估了来自同一制造商的维生素 E 注入聚乙烯衬垫(E1 和 E-poly, ZimmerBiomet,印第安纳州华沙)与中交联聚乙烯(ModXLPE)衬垫在初次 THA 中的翻修率。
患者与方法-我们基于芬兰关节置换登记处的数据进行了一项研究。研究组包括 2723 例使用维生素 E 注入衬垫的 THA 和 2707 例使用中交联聚乙烯衬垫的参考组。比较了两组之间的生存率、翻修风险和再翻修原因。
结果-维生素 E 注入聚乙烯衬垫组和以任何原因翻修为终点的参考组的 7 年生存率相当(分别为 94%[95%CI 92.9-94.9]和 93%[CI 91.9-93.9])。两组之间任何翻修的调整后的危险比(HR)相似(0.7[CI 0.4-1.1])。当以无菌性松动翻修为终点时,研究组的生存率为 99%(CI 98.6-99.4),参考组为 99%(CI 98.7-99.5),两组之间的翻修风险相似(HR 1.3[CI 0.7-2.5])。
结论-在 7 年的观察期后,维生素 E 注入衬垫的结果与交联聚乙烯衬垫的结果相当。