2,723 例维生素 E 注入的超高交联聚乙烯衬垫在全髋关节置换术中的植入存活率:来自芬兰关节置换登记处的数据。

Implant survival of 2,723 vitamin E-infused highly crosslinked polyethylene liners in total hip arthroplasty: data from the Finnish Arthroplasty Register.

机构信息

Department of Orthopaedic Surgery, University of Turku and Turku University Hospital, Turku.

Turku PET Centre, University of Turku and Turku University Hospital, Turku.

出版信息

Acta Orthop. 2021 Jun;92(3):316-322. doi: 10.1080/17453674.2021.1879513. Epub 2021 Feb 1.

Abstract

Background and purpose - The use of crosslinked polyethylene in total hip arthroplasty (THA) has decreased wear remarkably. It has been suggested that the antioxidative effects of vitamin E may enhance the wear properties of polyethylene even further. This study evaluates revision rates between vitamin E-infused polyethylene liners (E1 and E-poly, ZimmerBiomet, Warsaw, IN, USA) versus moderately crosslinked polyethylene (ModXLPE) liners from the same manufacturer used in primary THA.Patients and methods - We conducted a study based on data from the Finnish Arthroplasty Register. The study group consisted of 2,723 THAs with a vitamin E-infused liner and a reference group of 2,707 THAs with a moderately crosslinked polyethylene liner. Survivorship, revision risk, and re-revision causes were compared between groups.Results - The 7-year survival of the vitamin E-infused polyethylene liner group and of the reference group with revision for any reason as the endpoint was comparable (94% [95% CI 92.9-94.9] and 93% [CI 91.9-93.9], respectively). The adjusted hazard ratio (HR) for any revision was similar between the groups (0.7 [CI 0.4-1.1]). When revision for aseptic loosening was studied as the endpoint, the survival for the study group was 99% (CI 98.6-99.4) and for the reference group 99% (CI 98.7-99.5), and the risk of revision was comparable between the study groups (HR 1.3 [CI 0.7-2.5]).Interpretation - After an observation period of 7 years vitamin E-infused liners shows results equal to results obtained with crosslinked polyethylene liners.

摘要

背景与目的-在全髋关节置换术(THA)中使用交联聚乙烯可显著减少磨损。有人提出,维生素 E 的抗氧化作用可能会进一步提高聚乙烯的耐磨性能。本研究评估了来自同一制造商的维生素 E 注入聚乙烯衬垫(E1 和 E-poly, ZimmerBiomet,印第安纳州华沙)与中交联聚乙烯(ModXLPE)衬垫在初次 THA 中的翻修率。

患者与方法-我们基于芬兰关节置换登记处的数据进行了一项研究。研究组包括 2723 例使用维生素 E 注入衬垫的 THA 和 2707 例使用中交联聚乙烯衬垫的参考组。比较了两组之间的生存率、翻修风险和再翻修原因。

结果-维生素 E 注入聚乙烯衬垫组和以任何原因翻修为终点的参考组的 7 年生存率相当(分别为 94%[95%CI 92.9-94.9]和 93%[CI 91.9-93.9])。两组之间任何翻修的调整后的危险比(HR)相似(0.7[CI 0.4-1.1])。当以无菌性松动翻修为终点时,研究组的生存率为 99%(CI 98.6-99.4),参考组为 99%(CI 98.7-99.5),两组之间的翻修风险相似(HR 1.3[CI 0.7-2.5])。

结论-在 7 年的观察期后,维生素 E 注入衬垫的结果与交联聚乙烯衬垫的结果相当。