No differences in the body fat after violating core bioelectrical impedance measurement assumptions.

机构信息

Sherman Health Science Research Centre, School of Kinesiology and Health Science, York University, Rm 2002, 4700 Keele Street, Toronto, ON, M3J 1P3, Canada.

出版信息

BMC Public Health. 2021 Mar 12;21(1):495. doi: 10.1186/s12889-021-10552-y.

Abstract

OBJECTIVE

It is unclear to what degree acutely violating bioelectrical impedance analysis (BIA) measurement assumptions will alter the predicted percent fat mass (%FM) and whether this differs by sex or body mass index (BMI).

METHODS

%FM was assessed under control, dehydration, exercise, water, food intake and non-voided bladder conditions with three BIA devices (Tanita: BC-418, TBF-314, & Omron HBF-306CN) for men (n = 23, age: 24.0 ± 5.2 years old) and women (n = 17, age: 22.5 ± 3.4 years old) separately.

RESULTS

For all BIA devices, there were no differences in the %FM values between the control and the other conditions in men or women (- 1.9 to 0.4%, p > 0.05). Across the three devices and five conditions, 97% of %FM tests returned values within 5% of control (2 tests), and 86% of tests were within 2% of control despite violating an assumption. The errors were greatest with dehydration and women were more likely to have a %FM difference greater than 2% than men with dehydration using the hand-to-foot device (Tanita TBF-314: 59% versus 9%). There were no differences in %FM between control and the conditions when examined by BMI (overweight/obesity: - 2.8 to 0.1% and normal weight: - 1.7 to 0.5%; BMI*trial, p = 0.99).

CONCLUSION

%FM estimates were similar despite acutely violating the preliminary measurement BIA assumptions across a range of different BMIs. The minor variations in %FM are smaller than what would be expected with day-to-day variability or weight loss intervention but may be larger in women than men.

摘要

目的

目前尚不清楚在多大程度上违反生物电阻抗分析(BIA)测量假设会改变预测的体脂肪百分比(%FM),以及这种影响是否因性别或体重指数(BMI)而异。

方法

使用三种 BIA 设备(Tanita:BC-418、TBF-314 和 Omron HBF-306CN),在男性(n=23,年龄:24.0±5.2 岁)和女性(n=17,年龄:22.5±3.4 岁)中分别评估在控制、脱水、运动、水、食物摄入和未排空膀胱条件下的 %FM 值。

结果

对于所有 BIA 设备,男性或女性在控制和其他条件之间,%FM 值没有差异(-1.9%至 0.4%,p>0.05)。在三种设备和五种条件下,97%的%FM 测试值与对照值相差 5%以内(2 次测试),86%的测试值与对照值相差 2%以内,尽管违反了一个假设。脱水时的误差最大,脱水时女性的 %FM 差值大于 2%的可能性比男性高,使用手到脚设备(Tanita TBF-314:59%比 9%)。当按 BMI 检查时,%FM 在对照和条件之间没有差异(超重/肥胖:-2.8%至 0.1%和正常体重:-1.7%至 0.5%;BMI*trial,p=0.99)。

结论

尽管在不同 BMI 范围内急性违反 BIA 初步测量假设,但 %FM 估计值相似。%FM 的微小变化小于日常变异性或减肥干预所预期的变化,但在女性中可能比男性更大。