BMC Public Health. 2021 Mar 12;21(1):495. doi: 10.1186/s12889-021-10552-y.
It is unclear to what degree acutely violating bioelectrical impedance analysis (BIA) measurement assumptions will alter the predicted percent fat mass (%FM) and whether this differs by sex or body mass index (BMI).
%FM was assessed under control, dehydration, exercise, water, food intake and non-voided bladder conditions with three BIA devices (Tanita: BC-418, TBF-314, & Omron HBF-306CN) for men (n = 23, age: 24.0 ± 5.2 years old) and women (n = 17, age: 22.5 ± 3.4 years old) separately.
For all BIA devices, there were no differences in the %FM values between the control and the other conditions in men or women (- 1.9 to 0.4%, p > 0.05). Across the three devices and five conditions, 97% of %FM tests returned values within 5% of control (2 tests), and 86% of tests were within 2% of control despite violating an assumption. The errors were greatest with dehydration and women were more likely to have a %FM difference greater than 2% than men with dehydration using the hand-to-foot device (Tanita TBF-314: 59% versus 9%). There were no differences in %FM between control and the conditions when examined by BMI (overweight/obesity: - 2.8 to 0.1% and normal weight: - 1.7 to 0.5%; BMI*trial, p = 0.99).
%FM estimates were similar despite acutely violating the preliminary measurement BIA assumptions across a range of different BMIs. The minor variations in %FM are smaller than what would be expected with day-to-day variability or weight loss intervention but may be larger in women than men.
目前尚不清楚在多大程度上违反生物电阻抗分析(BIA)测量假设会改变预测的体脂肪百分比(%FM),以及这种影响是否因性别或体重指数(BMI)而异。
使用三种 BIA 设备(Tanita:BC-418、TBF-314 和 Omron HBF-306CN),在男性(n=23,年龄:24.0±5.2 岁)和女性(n=17,年龄:22.5±3.4 岁)中分别评估在控制、脱水、运动、水、食物摄入和未排空膀胱条件下的 %FM 值。
对于所有 BIA 设备,男性或女性在控制和其他条件之间,%FM 值没有差异(-1.9%至 0.4%,p>0.05)。在三种设备和五种条件下,97%的%FM 测试值与对照值相差 5%以内(2 次测试),86%的测试值与对照值相差 2%以内,尽管违反了一个假设。脱水时的误差最大,脱水时女性的 %FM 差值大于 2%的可能性比男性高,使用手到脚设备(Tanita TBF-314:59%比 9%)。当按 BMI 检查时,%FM 在对照和条件之间没有差异(超重/肥胖:-2.8%至 0.1%和正常体重:-1.7%至 0.5%;BMI*trial,p=0.99)。
尽管在不同 BMI 范围内急性违反 BIA 初步测量假设,但 %FM 估计值相似。%FM 的微小变化小于日常变异性或减肥干预所预期的变化,但在女性中可能比男性更大。