To compare central corneal thickness (CCT) measurement results obtained by RTVue OCT, Lenstar, Sirius topography, and ultrasound pachymetry (UP) (OcuScan RxP Ophthalmic Ultrasound System, Alcon Laboratories) in healthy subjects.
256 eyes of 128 healthy subjects were included in the study. CCT measurements were obtained from the eyes for each subject using the UP, Lenstar, Sirius topography, and RTVue OCT instruments. Ultrasound pachymetry measurements were performed after the eyes were anesthetized with one drop of 0.5% proparacaine hydrochloride and were performed at the same day after 15 minutes after previous measurements.
Of 128 participants, 84 were male and 44 were female with a mean age of 33.15 ± 12.95 years (ranging from 18 to 75 years). The mean CCTs for UP, Lenstar, Sirius topography, and RTVue OCT were 535.60 ± 35.15 (440,00-668,00); 532.63 ± 34.44 (449,00-650,00); 526.05 ± 36.45 (314,00-640,00); 525.89 ± 33.21 (437,00-646,00) µm, respectively. The mean CCT measurements were statistically different among the three groups (p = 0.002). CCT measurements obtained using RTVue-OCT were significantly thinner than those obtained using UP (p = 0.009). Sirius CCT measurements were also thinner than UP (p = 0.011). Mean CCT measurements between UP and Lenstar were similar (p = 0.769). Measurements with Lenstar (r = 0.849), Sirius topography (r = 0.883), and RTVue OCT (0.949) were highly correlated with UP measurements.
RTVue OCT and Sirius topography significantly underestimated the CCT compared with UP. Although highly correlated, the measurement values with these devices are not directly interchangeable in clinical practice.
比较RTVue光学相干断层扫描(OCT)、Lenstar、Sirius角膜地形图仪以及超声角膜测厚仪(UP,爱尔康实验室公司的OcuScan RxP眼科超声系统)在健康受试者中测量中央角膜厚度(CCT)的结果。
本研究纳入了128名健康受试者的256只眼。使用UP、Lenstar、Sirius角膜地形图仪和RTVue OCT仪器对每位受试者的眼睛进行CCT测量。在用一滴0.5%盐酸丙美卡因对眼睛进行麻醉后进行超声角膜测厚测量,且在前一次测量15分钟后的同一天进行。
128名参与者中,男性84名,女性44名,平均年龄为33.15±12.95岁(年龄范围为18至75岁)。UP、Lenstar、Sirius角膜地形图仪和RTVue OCT测量的平均CCT分别为535.60±35.15(440.00 - 668.00);532.63±34.44(449.00 - 650.00);526.05±36.45(314.00 - 640.00);525.89±33.21(437.00 - 646.00)µm。三组之间的平均CCT测量值在统计学上存在差异(p = 0.002)。使用RTVue - OCT获得的CCT测量值显著低于使用UP获得的值(p = 0.009)。Sirius测量的CCT值也低于UP测量值(p = 0.011)。UP和Lenstar之间的平均CCT测量值相似(p = 0.769)。Lenstar(r = 0.849)、Sirius角膜地形图仪(r = 0.883)和RTVue OCT(0.949)的测量值与UP测量值高度相关。