Department of Ophthalmology (H.G., İ.A., S. Kaya, S.S., A.C.Ç., K.S., V.G.), Trakya University of Medicine, Edirne, Turkey; and Department of Biostatistics and Medical Informatics (S. Korkmaz), Trakya University of Medicine, Edirne, Turkey.
Eye Contact Lens. 2021 Oct 1;47(10):539-545. doi: 10.1097/ICL.0000000000000787.
The aim of this study is to compare anterior segment parameters, including corneal thickness (CCT), keratometry and anterior chamber depth (ACD), and white to white corneal diameter (WTW), obtained by Pentacam Schiempflug imaging and intraocular lens (IOL) Master 700 swept-source optic coherence tomography biometry in keratoconus patients and healthy subjects.
This prospective cross-sectional instrument agreement analysis includes 88 eyes of 50 keratoconus patients and 87 eyes of 50 healthy subjects. Biometry was performed using IOL Master 700, and topography was performed using Pentacam. The keratometry values (Kf, Ks, Km, and Kmax), ACD, WTW, CCT, axial length (AL), anterior chamber angle (ACA), and lens thickness (LT) were evaluated. Levels of agreement between devices were evaluated by Bland-Altman plots with 95% limits of agreement.
Intraocular lens Master 700 showed higher WTW, ACD, pupil diameter, and CCT values than Pentacam in both the keratoconus and control groups. However, there were no statistically significant differences in flat keratometry (Kf) and steep keratometry (Ks) values between the groups.
Pentacam and IOL Master 700 may be used interchangeably in normal eyes and keratoconus eyes for the measurement of keratometry values and axis; however, these two devices should not be considered interchangeable for WTW, ACD, pupil diameter, and CCT measurements in both keratoconus patients and healthy subjects.
本研究旨在比较圆锥角膜患者和健康受试者的眼前节参数,包括角膜厚度(CCT)、角膜曲率(Keratometry)和前房深度(ACD)以及角膜直径(WTW),这些参数分别由 Pentacam Scheimpflug 成像和 IOL Master 700 扫频源光相干断层扫描生物测量仪获得。
本前瞻性横断面仪器一致性分析纳入了 50 例圆锥角膜患者的 88 只眼和 50 例健康受试者的 87 只眼。使用 IOL Master 700 进行生物测量,使用 Pentacam 进行地形图测量。评估角膜曲率值(Kf、Ks、Km 和 Kmax)、ACD、WTW、CCT、眼轴(AL)、前房角(ACA)和晶状体厚度(LT)。通过 Bland-Altman 图和 95%一致性界限评估两种设备之间的一致性水平。
在圆锥角膜组和对照组中,IOL Master 700 比 Pentacam 显示出更高的 WTW、ACD、瞳孔直径和 CCT 值。然而,两组之间平坦角膜曲率(Kf)和陡峭角膜曲率(Ks)值无统计学差异。
在正常眼和圆锥角膜眼中,Pentacam 和 IOL Master 700 可互换用于测量角膜曲率值和轴向;然而,在圆锥角膜患者和健康受试者中,这两种设备不应被视为 WTW、ACD、瞳孔直径和 CCT 测量的可互换设备。