Ma Shan, Li Cheng, Sun Jing, Yang Jun, Wen Kai, Chen Xi-Teng, Zhao Fang-Yu, Gao Rong-Yu, Tian Fang
Tianjin Key Laboratory of Retinal Functions and Diseases, Tianjin Branch of National Clinical Research Center for Ocular Disease, Eye Institute and School of Optometry, Tianjin Medical University Eye Hospital, Tianjin 300380, China.
Zhengda Guangming Eye Group, Weifang 261000, Shandong Province, China.
Int J Ophthalmol. 2024 Aug 18;17(8):1437-1446. doi: 10.18240/ijo.2024.08.08. eCollection 2024.
To assess and compare the variations and agreements across different ocular biometric parameters using swept-source optical coherence tomography (SS-OCT) and Scheimpflug tomography in patients diagnosed with cataract.
This prospective case series was conducted at Tianjin Medical University Eye Hospital. In total, 212 eyes from 212 patients scheduled for phacoemulsification were included. Eyes were evaluated preoperatively using two SS-OCT devices (IOLMaster700 and CASIA2) and Scheimpflug tomography (Pentacam). Central corneal thickness (CCT), anterior chamber depth (ACD), aqueous depth (AQD), white-to-white distance (WTW), flat simulated keratometry (Kf), steep simulated keratometry (Ks), mean keratometry (Km), and total corneal keratometry (TKm) were measured. Intraclass correlation coefficient (ICC), 95% confidence intervals (CI) and limits of agreement (LoA) widths were conducted to assess differences and correlations between devices.
All parameters, except for Ks, were significantly different. Pairwise comparison revealed no significant differences between keratometry obtained by IOLMaster 700 and Pentacam. LoA widths of all paired comparisons for Ks were >0.80 D. Except for WTW between IOLMaster 700 and CASIA2 and between CASIA2 and Pentacam, other Pearson's coefficients between devices showed a strong correlation (all >0.95). The ICC of WTW (ICC=0.438, 95%CI 0.167-0.625) showed poor reliability. The reliability of CCT, ACD, and AQD was excellent (all ICC>0.95), whereas that of TKm was good (ICC=0.827, 95%CI 0.221-0.939). A significant linear correlation was also observed among devices.
The ocular parameters derived from the use of IOLMaster700, CASIA2, and Pentacam exhibit significant discrepancies; as such, measurements from these devices should not be deemed as interchangeable.
使用扫频光学相干断层扫描(SS-OCT)和Scheimpflug断层扫描技术,评估并比较诊断为白内障患者的不同眼生物测量参数的差异及一致性。
本前瞻性病例系列研究在天津医科大学眼科医院进行。共纳入212例计划行超声乳化白内障吸除术患者的212只眼。术前使用两台SS-OCT设备(IOLMaster700和CASIA2)和Scheimpflug断层扫描(Pentacam)对眼睛进行评估。测量中央角膜厚度(CCT)、前房深度(ACD)、房水深度(AQD)、白对白距离(WTW)、模拟平角膜曲率(Kf)、模拟陡角膜曲率(Ks)、平均角膜曲率(Km)和总角膜曲率(TKm)。采用组内相关系数(ICC)、95%置信区间(CI)和一致性界限(LoA)宽度来评估不同设备之间的差异和相关性。
除Ks外,所有参数均存在显著差异。两两比较显示,IOLMaster 700和Pentacam测量的角膜曲率之间无显著差异。Ks所有配对比较的LoA宽度均>0.80 D。除IOLMaster 700与CASIA2之间以及CASIA2与Pentacam之间的WTW外,其他设备之间的Pearson系数显示出强相关性(均>0.95)。WTW的ICC(ICC = 0.438,95%CI 0.167 - 0.625)显示可靠性较差。CCT、ACD和AQD的可靠性极佳(所有ICC>0.95),而TKm的可靠性良好(ICC = 0.827,95%CI 0.221 - 0.939)。不同设备之间也观察到显著的线性相关性。
使用IOLMaster700、CASIA2和Pentacam得出的眼参数存在显著差异;因此,这些设备的测量结果不应被视为可互换的。