J Biomed Phys Eng. 2021 Oct 1;11(5):641-652. doi: 10.31661/jbpe.v0i0.1912-1017. eCollection 2021 Oct.
QT-interval prolongation is an important parameter for heart arrhythmia diagnosis. It is the time interval from QRS-onset to the T-end of electrocardiogram (ECG). Manual measurement of QT-interval, especially for 12-leads ECG, is time-consuming. Hence, an automatic QT-interval measurement is necessary. A new method for automatic QT-interval measurement is presented in this paper, which mainly consists of three parts, including QRS-complex detection, determination of QRS-onset, and T-end determination. The QRS-complex detection is based on the modified Pan-Tompkins algorithm. The T-end is defined based on Region of Interest (ROI) maximum limit. We compare and test our proposed QT-interval measurement method with reference measurement in term of correlation coefficient and range of 95% LoA. The correlation coefficient and the range of 95% LoA are 0.575 and 0.290, respectively. The proposed method is successfully implemented in ECG monitoring system using smartphone with high performance. The accuracy, positive predictive, and sensitivity of the QRS-complex detection in the system are 99.70%, 99.78%, and 99.92%, respectively. The range of 95% LoA for the comparison between manual and the system's QT-interval measurement is 0.216. The results show that the proposed method is dependable on the measure of the QT-interval and outperforms the other methods in term of correlation coefficient and range of 95% LoA.
QT间期延长是心律失常诊断的一个重要参数。它是心电图(ECG)从QRS波起始到T波结束的时间间隔。手动测量QT间期,尤其是对于12导联心电图,非常耗时。因此,自动测量QT间期是必要的。本文提出了一种自动测量QT间期的新方法,该方法主要由三部分组成,包括QRS波群检测、QRS波起始点确定和T波终点确定。QRS波群检测基于改进的Pan-Tompkins算法。T波终点基于感兴趣区域(ROI)最大限值来定义。我们将提出的QT间期测量方法与参考测量方法在相关系数和95%一致性界限(LoA)范围内进行比较和测试。相关系数和95% LoA范围分别为0.575和0.290。所提出的方法在高性能智能手机的心电图监测系统中成功实现。该系统中QRS波群检测的准确率、阳性预测值和灵敏度分别为99.70%、99.78%和99.92%。手动测量与系统QT间期测量之间比较的95% LoA范围为0.216。结果表明,所提出的方法在QT间期测量方面是可靠的,并且在相关系数和95% LoA范围方面优于其他方法。