Department of Urology, University Hospital of Patras, Rio, 26500, Patras, Greece.
Department of Urology, Medical University of Vienna, Vienna, Austria.
Urolithiasis. 2022 Oct;50(5):613-618. doi: 10.1007/s00240-022-01342-1. Epub 2022 Jun 30.
To compare the effect of MOSES™ modulation technology to conventional pulse delivery technology on the irrigation fluid temperature (IFT) under different irrigation conditions during flexible ureteroscopy (FURS) in a live-anesthetized porcine model. For this experiment was used one female pig. A percutaneous access was obtained and a 30Fr sheath was placed inside the upper calyceal system. A thermocouple was inserted through the sheath to the upper calyx to record the effect on IFT during FURS. A Lumenis 120H Ho:YAG laser was used and the IFT was recorded during laser activation for 30 s at a laser power of 20 W, 40 W and 60 W under gravity and manual pump irrigation using MOSES™ and conventional pulse delivery technology. In the highest power settings the maximum IFT was achieved in 18 s under gravity irrigation (66.4 °C). It seems that there is no significant difference on IFT between MOSES and conventional mode on the IFT under different irrigation conditions during FURS at 20 W, 40 W and 60 W power settings. Furthermore, our results indicate that under manual pumping even high-power settings (40 W, 60 W) can be performed with safety. In the in vivo model, the MOSES™ pulse delivery technology does not have a significant difference in the maximal IFT in comparison to conventional pulse delivery technology during FURS in the same power settings. Manual pumping should be used to keep the IFT within safe limits.
为了比较 MOSES™ 调制技术与传统脉冲输送技术在活体麻醉猪模型下经软性输尿管镜检查术(FURS)中不同冲洗条件下对冲洗液温度(IFT)的影响。该实验使用了一只雌性猪。经皮入路,在上盏系统内置入 30Fr 鞘。将热电偶插入鞘内至上盏,以记录 FURS 过程中对 IFT 的影响。使用 Lumenis 120H Ho:YAG 激光,在重力和手动泵冲洗下,以 20W、40W 和 60W 的激光功率记录激光激活 30s 期间的 IFT,采用 MOSES™ 和传统脉冲输送技术。在最高功率设置下,重力冲洗时达到最大 IFT(66.4°C)的时间为 18s。在 FURS 期间,20W、40W 和 60W 的功率设置下,MOSES 和传统模式下的 IFT 似乎没有显著差异。此外,我们的结果表明,在手动泵冲洗下,即使在高功率设置(40W、60W)下也可以安全进行。在体内模型中,在相同的功率设置下,MOSES™ 脉冲输送技术与传统脉冲输送技术相比,最大 IFT 无显著差异。应使用手动泵以将 IFT 保持在安全范围内。