Nachman Dean, Eisenkraft Arik, Goldstein Nir, Ben-Ishay Arik, Fons Meir, Merin Roei, Gepner Yftach
Heart Institute, Hadassah Ein Kerem Medical Center, Jerusalem, Israel.
Institute for Research in Military Medicine, Faculty of Medicine, The Hebrew University of Jerusalem and the Israel Defense Force Medical Corps, Jerusalem, Israel.
Front Physiol. 2022 Jun 29;13:911544. doi: 10.3389/fphys.2022.911544. eCollection 2022.
Vital signs obtained by photoplethysmography-based devices might be influenced by subcutaneous fat and skin color. This observational comparison study aimed to test the accuracy of blood pressure (BP) measurements between a photoplethysmography-based device and cuff-based BP device in ambulatory individuals, coming for a routine BP checkup. Systolic BP (SBP) and diastolic BP (DBP) measurements were stratified based on sex, BMI (<25; 25 ≤BMI<30; 30 ≤kg/m), and skin color (types 1-3 and 4-6 by the Fitzpatrick scale). A total of 1548 measurements were analyzed. Correlations of SBP and DBP between the devices among males/females were between 0.914-0.987 ( < 0.001), and Bland-Altman analysis showed a bias of less than 0.5 mmHg for both sexes. Correlations of SBP and DBP between the devices among BMI groups were between 0.931-0.991 ( < 0.001), and Bland-Altman analysis showed a bias of less than 1 mmHg for all. Correlations of SBP and DBP between the devices among the skin color groups were between 0.936-0.983 ( < 0.001), and Bland-Altman analysis showed a bias of less than 1 mmHg for all. This study shows similar and high agreements between BP measurements obtained using a PPG-based non-invasive cuffless BP device and a cuff-based BP device across sex, BMI, and skin color groups.
基于光电容积脉搏波描记法的设备所获取的生命体征可能会受到皮下脂肪和肤色的影响。这项观察性比较研究旨在测试基于光电容积脉搏波描记法的设备与基于袖带的血压设备在前来进行常规血压检查的非卧床个体中测量血压(BP)的准确性。收缩压(SBP)和舒张压(DBP)测量值根据性别、体重指数(BMI,<25;25≤BMI<30;30≤kg/m)和肤色(根据菲茨帕特里克量表分为1 - 3型和4 - 6型)进行分层。总共分析了1548次测量数据。男性/女性中两种设备之间SBP和DBP的相关性在0.914 - 0.987之间(<0.001),布兰德 - 奥特曼分析显示两性的偏差均小于0.5 mmHg。BMI组中两种设备之间SBP和DBP的相关性在0.931 - 0.991之间(<0.001),布兰德 - 奥特曼分析显示所有组的偏差均小于1 mmHg。肤色组中两种设备之间SBP和DBP的相关性在0.936 - 0.983之间(<0.001),布兰德 - 奥特曼分析显示所有组的偏差均小于1 mmHg。这项研究表明,在不同性别、BMI和肤色组中,使用基于光电容积脉搏波描记法的无创无袖带血压设备和基于袖带的血压设备所获得的血压测量值之间具有相似且高度的一致性。