Woolfrey B F, Lally R T, Ederer M N
J Clin Microbiol. 1986 Mar;23(3):629-30. doi: 10.1128/jcm.23.3.629-630.1986.
A total of 398 consecutive clinical staphylococcus isolates, of which 205 were coagulase negative and 193 were coagulase positive, were tested in parallel by using AutoMicrobic system Gram-Positive Susceptibility-MIC cards and modified Mueller-Hinton agar containing 4% NaCl and oxacillin (6 micrograms/ml). The AutoMicrobic system cards correctly detected 103 of 104 (99%) oxacillin-resistant coagulase-negative isolates with no reports of false resistance.
总共398株连续的临床葡萄球菌分离株,其中205株凝固酶阴性,193株凝固酶阳性,使用自动微生物系统革兰氏阳性药敏-MIC卡和含4%氯化钠及苯唑西林(6微克/毫升)的改良穆勒-欣顿琼脂进行平行检测。自动微生物系统卡正确检测出104株耐苯唑西林凝固酶阴性分离株中的103株(99%),无假耐药报告。