Harper J A, Gal A, Burchell R K, Summers J D, Starling J, Gerber K, Gummow B
North Coast Specialist and Referral Centre, 5/30 Chancellor Village, Sippy Downs, Sunshine Coast, Australia.
Department of Veterinary Clinical Medicine, College of Veterinary Medicine, Carle Illinois College of Medicine, University of Illinois, Urbana, Illinois, USA.
J Small Anim Pract. 2023 May;64(5):337-342. doi: 10.1111/jsap.13588. Epub 2023 Feb 2.
To compare the ionised calcium measured on a portable analyser (iSTAT, Abbott) to a reference method.
Blood samples from 39 apparently healthy dogs were analysed in duplicate using a portable analyser and a reference method (Radiometer ABL800 FLEX). Bland-Altman plots and Passing-Bablok regression were used to assess constant and proportional bias between the two instruments. A within-assay percentage coefficient of variation and total error (TE) was calculated for both analysers. The reference interval was calculated for the portable analyser using the robust method with confidence interval bootstrapping.
The Bland-Altman plot showed a -0.036 mmol/L difference between the two instruments (95% confidence limit -0.08 to 0.01 mmol/L; limits of agreement -0.07 to 0.006 mmol/L). Neither the Bland-Altman plot nor the Passing-Bablock regression (slope -0.03; 95% confidence interval -0.08 to 0.19 and intercept 1; 95% confidence interval 0.83 to 1.2) showed significant proportional bias. The coefficient of variation for the portable analyser was 1.08%, compared to 0.78% for the reference method with a total error of 3.5% for the portable analyser. The estimated population-based reference interval for ionised calcium using the portable analyser is 1.23 to 1.42 mmol/L.
For the healthy dogs in this study, compared to the reference method, the portable analyser showed no significant bias for measurement of ionised calcium. Further studies including hyper and hypocalcaemic dogs are required to determine clinical impact of the use of this analyser.
比较便携式分析仪(iSTAT,雅培)和参考方法测量的离子钙。
使用便携式分析仪和参考方法(Radiometer ABL800 FLEX)对 39 只健康犬的血液样本进行重复分析。使用 Bland-Altman 图和 Passing-Bablok 回归来评估两种仪器之间的恒偏和比例偏倚。为两种分析仪计算了批内变异系数和总误差(TE)。使用置信区间自举的稳健方法为便携式分析仪计算参考区间。
Bland-Altman 图显示两种仪器之间存在-0.036mmol/L 的差异(95%置信限-0.08 至 0.01mmol/L;允许区间-0.07 至 0.006mmol/L)。Bland-Altman 图和 Passing-Bablock 回归(斜率-0.03;95%置信区间-0.08 至 0.19 和截距 1;95%置信区间 0.83 至 1.2)均未显示出明显的比例偏倚。便携式分析仪的变异系数为 1.08%,参考方法为 0.78%,便携式分析仪的总误差为 3.5%。使用便携式分析仪估计的健康犬离子钙的人群参考区间为 1.23 至 1.42mmol/L。
与参考方法相比,在这项研究中,对于健康犬,便携式分析仪在测量离子钙方面没有显示出显著的偏差。需要进一步研究包括高钙血症和低钙血症的犬,以确定使用该分析仪的临床影响。