Suppr超能文献

OrbscanIIz与超声测厚法测量中央角膜厚度的比较:常规临床实践中的一致性研究

Comparison of central corneal thickness measurements obtained by OrbscanIIz and ultrasound pachymetry: A concordance study in the usual clinical practice.

作者信息

Romero-Trevejo José L, Sánchez-Pérez Andrés, Muñoz-García Erika, Fernández-Romero Lourdes, Jiménez-Navarro Manuel F

机构信息

Department of Ophthalmology. Virgen de la Victoria University Hospital, Malaga, Spain.

Department of Medicine and Dermatology, School of Medicine, University of Malaga, Malaga, Spain.

出版信息

Saudi J Ophthalmol. 2023 Mar 9;37(1):10-14. doi: 10.4103/sjopt.Sjopt_82_21. eCollection 2023 Jan-Mar.

Abstract

PURPOSE

The objective of this study was to assess the concordance between the values obtained in measuring central corneal thickness using the OrbscanIIz and the contact ultrasonic pachymeter available in our public ophthalmology service.

METHODS

Measurements were taken from 88 eyes of 44 patients using the two instruments. The data obtained were statistically analyzed using version 22 of the IBM SPSS program.

RESULTS

The mean of central corneal thickness measurements obtained from OrbscanIIz was significantly higher than that obtained from ultrasound pachymetry. However, the mean of differences between both instruments was only 7.22 μ, which could be considered a clinically insignificant result when considering the good concordance obtained between both systems.

CONCLUSION

OrbscanIIz and ultrasound pachymetry can be interchangeable in the usual public clinical practice when measuring central corneal thickness. This is the first research found in the literature that uses a concordance study to compare the data resulting from central corneal thickness measurements obtained by an OrbscanIIz and an OcuScan pachymeter in our environment.

摘要

目的

本研究的目的是评估使用OrbscanIIz测量中央角膜厚度所获得的值与我们公共眼科服务中可用的接触式超声角膜测厚仪所获得的值之间的一致性。

方法

使用这两种仪器对44例患者的88只眼睛进行测量。使用IBM SPSS程序22版对获得的数据进行统计分析。

结果

OrbscanIIz获得的中央角膜厚度测量平均值显著高于超声测厚法获得的平均值。然而,两种仪器之间的平均差值仅为7.22μm,考虑到两个系统之间获得的良好一致性,这可以被认为是临床上无显著意义的结果。

结论

在测量中央角膜厚度时,OrbscanIIz和超声测厚法在普通公共临床实践中可以互换使用。这是文献中首次发现的使用一致性研究来比较在我们环境中由OrbscanIIz和OcuScan角膜测厚仪获得的中央角膜厚度测量数据的研究。

https://cdn.ncbi.nlm.nih.gov/pmc/blobs/35e8/10032292/a2a512438cf9/SJO-37-10-g001.jpg

文献AI研究员

20分钟写一篇综述,助力文献阅读效率提升50倍。

立即体验

用中文搜PubMed

大模型驱动的PubMed中文搜索引擎

马上搜索

文档翻译

学术文献翻译模型,支持多种主流文档格式。

立即体验