Davies G G, Jebson P J, Glasgow B M, Hess D R
Crit Care Med. 1986 Oct;14(10):881-5. doi: 10.1097/00003246-198610000-00010.
A system has been developed to monitor continuously the components of the oxygen Fick equation: oxygen consumption by a gas exchange analyzer and arteriovenous oxygen difference by pulse and fiberoptic oximetry. A computer-based system was developed which calculates cardiac output and other variables every 20 sec. Continuous Fick (CF) cardiac output was compared to thermodilution (TD) cardiac output in 21 ventilated post-cardiac surgery patients. A total of 237 simultaneous cardiac output measurements had a range between 2 and 11 L/min. The correlation between CF and TD cardiac outputs was r = .86, with an equation of TD cardiac output = 0.92 CF cardiac output + 1.16. There was a significant (p less than .001) difference between the two methods of cardiac output estimation. The CF method was consistently lower than TD; this difference was greater at lower flows. CF cardiac output measurement is practical; it offers distinct advantages in viewing cardiac output together with oxygen demand and oxygen extraction.
已开发出一种系统,可连续监测氧耗量菲克方程的各个组成部分:通过气体交换分析仪监测氧耗量,通过脉搏和光纤血氧测定法监测动静脉氧分压差。开发了一种基于计算机的系统,该系统每20秒计算一次心输出量和其他变量。对21例心脏手术后接受机械通气的患者,将连续菲克(CF)心输出量与热稀释(TD)心输出量进行了比较。总共进行了237次同步心输出量测量,范围在2至11升/分钟之间。CF心输出量与TD心输出量之间的相关性为r = 0.86,TD心输出量的方程为TD心输出量= 0.92×CF心输出量+ 1.16。两种心输出量估计方法之间存在显著差异(p < 0.001)。CF方法始终低于TD;在较低流量时这种差异更大。CF心输出量测量是切实可行的;它在将心输出量与氧需求和氧摄取量一起观察方面具有明显优势。