Saminathan Sathiyan, Kannan Mageshraja, Ganesh K M, Mazarello James, Fernandes Rahul
Department of Radiation Physics, Kidwai Memorial Institute of Oncology, Bengaluru, Karnataka, India.
Roaslina India Private Limited, Mumbai, Maharashtra, India.
J Med Phys. 2023 Oct-Dec;48(4):333-337. doi: 10.4103/jmp.jmp_76_23. Epub 2023 Dec 5.
To investigate the dosimetric performance of newly developed parallel plate chamber in electron beams.
Rosalina Instruments India Private Limited (Mumbai, Maharashtra, India) has designed and fabricated PRATT2 parallel plate chamber. The various dosimetric characteristics, including pre- and post-irradiation leakage, stability, polarity effect, chamber response with bias voltage, dose linearity, dose rate effect, and chamber absorbed dose calibration, were performed for the developed chamber. The electron beam energies of 4, 6, 8, and 15MeV were used in this study.
The pre- and post-irradiation leakage of the developed chamber was within the acceptable limit. The chamber shows good stability over the electron beams used in this study. The maximum error in polarity effect was 0.7% for the developed chamber. The chamber shows the good linear response with dose, and its response is independent of the dose rate for all electron beams. The beam quality correction factor (k) was determined for the all electron beam energies, which was used for determination absorbed dose in electron beams.
The developed parallel plate chamber (PRATT2) is suitable for dosimetry of electron beams in radiotherapy. The chamber is cost effective and shows precise and reproducible response. The study carried out confirms that the newly designed and fabricated ion chamber can be used in the measurement of absorbed dose for therapeutic electron beams.
研究新开发的平行板电离室在电子束中的剂量学性能。
印度罗莎琳娜仪器私人有限公司(印度马哈拉施特拉邦孟买)设计并制造了PRATT2平行板电离室。对所开发的电离室进行了各种剂量学特性研究,包括辐照前后的泄漏、稳定性、极性效应、电离室在偏置电压下的响应、剂量线性、剂量率效应以及电离室吸收剂量校准。本研究使用了4、6、8和15MeV的电子束能量。
所开发电离室的辐照前后泄漏在可接受范围内。该电离室在本研究使用的电子束中表现出良好的稳定性。所开发电离室的极性效应最大误差为0.7%。该电离室对剂量显示出良好的线性响应,并且其响应与所有电子束的剂量率无关。确定了所有电子束能量的射束质量校正因子(k),用于确定电子束中的吸收剂量。
所开发的平行板电离室(PRATT2)适用于放射治疗中电子束的剂量测定。该电离室具有成本效益,并且显示出精确且可重复的响应。所进行的研究证实,新设计和制造的电离室可用于治疗性电子束吸收剂量的测量。