Burns D T, McEwen M R
Centre for Ionizing Radiation Metrology, National Physical Laboratory, Middlesex, UK.
Phys Med Biol. 1998 Aug;43(8):2033-45. doi: 10.1088/0031-9155/43/8/003.
The NACP electron chamber is one of three parallel-plate chambers recommended for use in the UK. Measurements with this chamber type have indicated a problem in determining the recombination correction. This is due to a variation of the ionization current I with polarizing voltage V which deviates from the accepted Boag theory. It is shown that there is a chamber-dependent threshold voltage below which the NACP chamber follows the Boag theory. Above this voltage the chamber should be used with caution, although it is still possible to correct for the dependence of the chamber response on the dose per pulse. The existence of such deviations from theory demonstrates the usefulness of the 1/I against 1/V plot and the limitations of the Boag two-voltage analysis. Values for the initial recombination and the coefficient of general recombination are measured for several NACP chambers. It is shown that from these one can derive a value for the effective plate separation and the collector radius of each chamber. Differences in the behaviour of NACP chambers manufactured by Scanditronix and Dosetek are discussed and the implications of free-electron collection are considered.
NACP电子室是英国推荐使用的三种平行板室之一。使用这种类型的室进行测量表明,在确定复合校正方面存在问题。这是由于电离电流I随极化电压V的变化偏离了公认的博阿格理论。结果表明,存在一个与室相关的阈值电压,低于该电压时,NACP室遵循博阿格理论。高于此电压时,应谨慎使用该室,尽管仍然可以校正室响应与每脉冲剂量的相关性。这种与理论的偏差表明了1/I对1/V曲线的有用性以及博阿格双电压分析的局限性。测量了几个NACP室的初始复合值和一般复合系数。结果表明,由此可以得出每个室的有效极板间距和收集器半径的值。讨论了Scanditronix和Dosetek生产的NACP室行为的差异,并考虑了自由电子收集的影响。