Brown W J
J Clin Microbiol. 1986 Mar;23(3):640-2. doi: 10.1128/jcm.23.3.640-642.1986.
A new commercial yellow latex agglutination reagent (Bacto-Staph) was compared with the slide and tube coagulase tests and three other commercial reagents for the identification of 283 Staphylococcus aureus and 54 non-S. aureus staphylococcal strains. Test sensitivities for the identification of S. aureus were as follows: tube coagulase, 99.6%; slide coagulase, 98.6%; Bacto-Staph, 99.6%; Staphylatex, 98.6%; Sero STAT Staph, 98.2%; and Staphyloslide, 97.5%. No false-positive reactions were observed with any of the commercial reagents.
一种新型商用黄色乳胶凝集试剂(Bacto-Staph)与玻片法和试管法凝固酶试验以及其他三种商用试剂进行了比较,用于鉴定283株金黄色葡萄球菌和54株非金黄色葡萄球菌葡萄球菌菌株。鉴定金黄色葡萄球菌的试验敏感性如下:试管法凝固酶,99.6%;玻片法凝固酶,98.6%;Bacto-Staph,99.6%;Staphylatex,98.6%;Sero STAT Staph,98.2%;以及Staphyloslide,97.5%。任何一种商用试剂均未观察到假阳性反应。