Sceptor微量稀释法抗生素药敏试验系统的评估:一项合作研究。

Evaluation of the Sceptor microdilution antibiotic susceptibility testing system: a collaborative investigation.

作者信息

Jones R N, Thornsberry C, Barry A L, Gavan T L

出版信息

J Clin Microbiol. 1981 Jan;13(1):184-94. doi: 10.1128/jcm.13.1.184-194.1981.

Abstract

A multiwell, dried antimicrobial agent susceptibility test system, Sceptor (BBL Microbiology Systems, Cockeysville, Md.), was tested. The system was compared directly with a reference microdilution method by using two collections of stock cultures and 305 fresh clinical isolates. Sceptor was found to be in agreement (+/- log2 dilution) with the reference microdilution method in 96.9 to 98.3% of 9,840 minimal inhibitory concentration determinations performed on stock strains and 95.0% of 7,308 minimal inhibitory concentrations obtained from the clinical isolates. The intralaboratory and interlaboratory reproducibility on stock strains was 97.6 and 97.2%, respectively. The intralaboratory reproducibility for the clinical isolates was 96.9%. Sceptor accurately categorized representative challenge strains of methicillin-resistant staphylococci, beta-lactamase-producing bacteria, organisms producing other antimicrobial agent-inactivating enzymes, and permeability mutants as resistant. Only 0.2% very major errors (false-sensitive minimal inhibitory concentrations by Sceptor) were identified among the clinical isolate test results, the majority being clinically insignificant. The product is accurate and reliable, has a long shelf life, and seems applicable for routine use in clinical laboratories.

摘要

对一种多孔干抗微生物药敏试验系统Sceptor(BBL微生物系统公司,马里兰州科基斯维尔)进行了测试。使用两组保藏培养物和305株新鲜临床分离株,将该系统与参考微量稀释法直接进行比较。在对保藏菌株进行的9840次最低抑菌浓度测定中,96.9%至98.3%的结果与参考微量稀释法一致(±log2稀释度);从临床分离株获得的7308个最低抑菌浓度中,95.0%的结果与参考微量稀释法一致。保藏菌株的实验室内和实验室间重复性分别为97.6%和97.2%。临床分离株的实验室内重复性为96.9%。Sceptor能准确地将耐甲氧西林葡萄球菌、产β-内酰胺酶细菌、产生其他抗微生物药物失活酶的微生物以及通透性突变体等代表性挑战菌株归类为耐药。在临床分离株测试结果中,仅发现0.2%的极重大错误(Sceptor得出的假敏感最低抑菌浓度),大多数在临床上无意义。该产品准确可靠,保质期长,似乎适用于临床实验室的常规使用。

相似文献

引用本文的文献

本文引用的文献

7
Evaluation of the performance parameters of a prediluted, quantitative antibiotic susceptibility test device.
Antimicrob Agents Chemother. 1977 Feb;11(2):271-6. doi: 10.1128/AAC.11.2.271.

文献AI研究员

20分钟写一篇综述,助力文献阅读效率提升50倍。

立即体验

用中文搜PubMed

大模型驱动的PubMed中文搜索引擎

马上搜索

文档翻译

学术文献翻译模型,支持多种主流文档格式。

立即体验