Meier Patricia, Calisti Maité, Werner Inge, Debertin Daniel, Mayer-Suess Lukas, Knoflach Michael, Federolf Peter
Department of Neurology, Medical University of Innsbruck, 6020 Innsbruck, Austria.
VASCage, Centre on Clinical Stroke Research, 6020 Innsbruck, Austria.
Sensors (Basel). 2025 Feb 21;25(5):1309. doi: 10.3390/s25051309.
In view of the frequent use of the portable balance board (Tymo, Tyromotion, Graz, Austria) in neurorehabilitation, the objective of this study is to assess its posturographic validity and reliability in comparison to a floor-embedded force platform (AMTI, Watertown, MA, USA), as well as its clinical validity in stroke patients. To evaluate posturographic validity, thirty healthy participants were tested in four different measurement conditions (M1-M4: standing with/without a balance pad, eyes open/closed) while simultaneously measuring with both devices placed on top of each other. Clinical validity is obtained by correlating stroke patients' Tymo outcome variables (path length and ellipse) of M1-M4 with Berg Balance Scale (BBS) results. Reliability is determined by correlating repeated measurements. The Center of pressure (COP) paths of both force plates show a very strong correlation. The repeated measurement reliability of both force plates is comparable. Tymo outcome variables exhibit moderate to strong correlation with the BBS and demonstrated repeated measurement reliability analogous to healthy participants. Therefore, the Tymo device demonstrates measurement accuracy comparable to a gold-standard device while offering the advantages of portability and user-friendly software. Tymo is suitable for assessing balance in stroke patients, though protocol adaptations and averaging repeated measurements are recommended to enhance feasibility and reliability.
鉴于便携式平衡板(Tymo,Tyromotion,奥地利格拉茨)在神经康复中经常使用,本研究的目的是评估其与地面嵌入式测力平台(AMTI,美国马萨诸塞州沃特敦)相比的姿势描记有效性和可靠性,以及其在中风患者中的临床有效性。为了评估姿势描记有效性,30名健康参与者在四种不同的测量条件下(M1 - M4:站立时有/无平衡垫,睁眼/闭眼)进行测试,同时将两种设备叠放在一起进行测量。通过将中风患者M1 - M4的Tymo结果变量(路径长度和椭圆)与伯格平衡量表(BBS)结果相关联来获得临床有效性。通过将重复测量结果相关联来确定可靠性。两个测力板的压力中心(COP)路径显示出非常强的相关性。两个测力板的重复测量可靠性相当。Tymo结果变量与BBS表现出中度至强相关性,并显示出与健康参与者类似的重复测量可靠性。因此,Tymo设备显示出与金标准设备相当的测量准确性,同时具有便携性和用户友好软件的优点。Tymo适用于评估中风患者的平衡,不过建议调整方案并对重复测量进行平均以提高可行性和可靠性。