Suppr超能文献

一款即时检验血气分析仪——诺瓦Stat Profile Primer Plus分析仪的验证过程

The Verification Process of a POC Blood Gas Analyser-The Nova Stat Profile Primer Plus Analyser.

作者信息

Costa Joseph, Attard Ritienne

机构信息

GGH Medical Laboratory Services, Gozo General Hospital, Victoria, Gozo, Malta.

出版信息

J Clin Lab Anal. 2025 Apr;39(7):e70006. doi: 10.1002/jcla.70006. Epub 2025 Mar 25.

Abstract

BACKGROUND

Point-of-care testing of blood gases plays a critical role in patient management. The aim of this study was to verify the manufacturer's specifications of the Nova Stat Profile Prime Plus Analyser, along with a comparison study with the GEM Premier 4000 Blood Gas Analyser.

METHODS

Parameters analysed were pH, pCO, pO, Na, Cl, K, iCa, lactate, and glucose. Data for the precision and bias study were generated using control samples in a 5 × 5 study design. Linearity was checked using a five-level Linearity Control Set, while comparison was done between the Nova and GEM analysers using whole blood samples (N = 103). Acceptance was based on the CLIA TE for all analytes except for lactate, for which the TE defined by CAP and AAB was used.

RESULTS

The within-run and between-run CV% precision were all lower than the claimed CVs%, except for pCO control level 2 within run (CV% 1.5 [claim CV% 1.1]) and iCa control level 5 between run (CV% 1.42 [claim 1.12]). The observed bias for all parameters was within the calculated lower and upper bias limits. Linearity was verified for all analytes except for Na. Upon comparison of the Nova and GEM analysers, a correlation coefficient above 0.95 was observed for most parameters.

CONCLUSION

The Nova Stat Profile Prime Plus analyser meets the manufacturer's precision and bias claims. Linearity was confirmed for most analytes. There was a good correlation between the Nova and GEM Blood Gas analyser at concentrations within the reference range intervals for all investigated parameters.

摘要

背景

血气即时检测在患者管理中起着关键作用。本研究的目的是验证诺瓦Stat Profile Prime Plus分析仪的制造商规格,并与GEM Premier 4000血气分析仪进行比较研究。

方法

分析的参数包括pH、pCO、pO、Na、Cl、K、iCa、乳酸和葡萄糖。使用5×5研究设计中的对照样本生成精密度和偏差研究数据。使用五级线性控制集检查线性,同时使用全血样本(N = 103)在诺瓦和GEM分析仪之间进行比较。除乳酸外,所有分析物的验收均基于CLIA TE,乳酸的TE使用CAP和AAB定义的TE。

结果

批内和批间CV%精密度均低于声称的CV%,除批内pCO对照水平2(CV% 1.5 [声称CV% 1.1])和批间iCa对照水平5(CV% 1.42 [声称1.12])。所有参数的观察偏差均在计算的上下偏差限内。除Na外,所有分析物的线性均得到验证。在比较诺瓦和GEM分析仪时,大多数参数的相关系数均高于0.95。

结论

诺瓦Stat Profile Prime Plus分析仪符合制造商的精密度和偏差要求。大多数分析物的线性得到确认。在所有研究参数的参考区间浓度范围内,诺瓦和GEM血气分析仪之间具有良好的相关性。

https://cdn.ncbi.nlm.nih.gov/pmc/blobs/eeae/11981951/a60ae742d3f1/JCLA-39-e70006-g002.jpg

文献检索

告别复杂PubMed语法,用中文像聊天一样搜索,搜遍4000万医学文献。AI智能推荐,让科研检索更轻松。

立即免费搜索

文件翻译

保留排版,准确专业,支持PDF/Word/PPT等文件格式,支持 12+语言互译。

免费翻译文档

深度研究

AI帮你快速写综述,25分钟生成高质量综述,智能提取关键信息,辅助科研写作。

立即免费体验