Thomas Linya, Kulkarni Chidanand, Kuzhuppilly Neetha I R
Department of Ophthalmology, Kasturba Medical College Manipal, Manipal Academy of Higher Education, Manipal, Karnataka, India.
Taiwan J Ophthalmol. 2024 Nov 8;15(3):466-473. doi: 10.4103/tjo.TJO-D-24-00058. eCollection 2025 Jul-Sep.
Optic disc size measurement is essential for determining landmarks, lesions, and distances of the retina. Due to the size variability in nature and when measured with different instruments, newer methods to compare sizes between different instruments are needed. We aimed to determine the comparability of slit lamp, fundus photography, and optical coherence tomography (OCT) measurements of optic disc size after correcting for both ocular and instrument magnification.
In this prospective, observational study, optic disc size was measured by slit-lamp biomicroscopy, fundus photography, and OCT. Instrument and ocular magnification was calculated, and the measured size was corrected for both using the Littman formula. The corrected values were analyzed using Bland-Altman plots and intra-class correlation (ICC) for agreeability and correlation, respectively. Linear regression analysis was conducted to estimate the systematic errors and interconvertibility in cases of high correlation.
Fifty-six eyes were included for the analysis. Uncorrected and corrected values for each method had excellent correlation and agreeability (ICC absolute agreement >0.75, < 0.01). Among magnification corrected values, slit lamp and OCT had excellent correlation and agreement (ICC consistency = 0.846, confidence interval 0.75-0.91, < 0.01). This pair had = 0.73 on linear regression ( < 0.01). Linear regression formulas for interconverting uncorrected and corrected values for the three methods had variable accuracy ( from 0.47 to 0.66).
Uncorrected values of different methods for disc size measurement cannot be compared directly; they need ocular and instrument magnification correction. Among the three methods compared, the best interconvertible pairs of measurement were magnification corrected slit lamp and OCT values.
视盘大小测量对于确定视网膜的标志、病变及距离至关重要。由于视盘大小在自然状态下存在变异性,且使用不同仪器测量时也会有差异,因此需要更新的方法来比较不同仪器测量的视盘大小。我们旨在确定在校正眼和仪器放大倍数后,裂隙灯、眼底照相及光学相干断层扫描(OCT)对视盘大小测量结果的可比性。
在这项前瞻性观察研究中,通过裂隙灯生物显微镜检查、眼底照相及OCT对视盘大小进行测量。计算仪器和眼的放大倍数,并使用利特曼公式对测量大小进行校正。分别使用布兰德-奥特曼图和组内相关系数(ICC)对校正后的值进行一致性和相关性分析。进行线性回归分析以估计高相关性情况下的系统误差和相互转换性。
纳入56只眼进行分析。每种方法的未校正值和校正值具有极好的相关性和一致性(ICC绝对一致性>0.75,P<0.01)。在放大倍数校正值中,裂隙灯和OCT具有极好的相关性和一致性(ICC一致性 = 0.846,置信区间0.75 - 0.91,P<0.01)。这一对在直线回归分析中的斜率为0.73(P<0.01)。三种方法未校正值和校正值相互转换的线性回归公式准确性各异(R²从0.47至0.66)。
不同方法测量视盘大小的未校正值不能直接比较;它们需要进行眼和仪器放大倍数校正。在比较的三种方法中,测量结果相互转换最佳的配对是放大倍数校正后的裂隙灯和OCT值。