UNLABELLED

A new method for the measurement of renin in human plasma is described. The method is based on the introduction of the internationally available renin standard of the Medical Research Council (MRC) London, as a calibration system. Thus, some principal disadvantages of methods expressing results in renin reaction velocity (angiotensin generation rate) only are avoided. Both renins, unknown and standard, react with a sheep substrate preparation and are handled identically throughout the whole procedure including the angiotensin I radioimmunoassay (RIA). The plasma renin concentration (PRC) is given in 10(-6) MRC-renin units (muM/ml).

RESULTS

the renin standard is free of angiotensin, angiotensinases, and angiotensinogen; it is stable on storage. Identical enzyme kinetics are shown for both renins. An interference between endogenous and exogenous substrate could be avoided. The potentially harmful influences of proteins from the enzyme incubation mixture of the RIA dose response curve are shown. The use of an angiotensin I calibration system could be omitted. Using a standard renin dilution from 250-0.9 muU/ml also the full biological range is covered. When giving an unrestricted diet the preliminary normal values of PRC are 21.9 +/- 12.6 muU/ml in recumbent and 40.1 +/- 19.8 muU/ml in upright position (n = 16,x +/- s, age 20-35 years). Earlier findings of age-dependency of PRC were confirmed.

摘要

文献检索

告别复杂PubMed语法,用中文像聊天一样搜索,搜遍4000万医学文献。AI智能推荐,让科研检索更轻松。

立即免费搜索

文件翻译

保留排版,准确专业,支持PDF/Word/PPT等文件格式,支持 12+语言互译。

免费翻译文档

深度研究

AI帮你快速写综述,25分钟生成高质量综述,智能提取关键信息,辅助科研写作。

立即免费体验